MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/brs.0000000000004652

Медицинская библиотека · Spine · 2023

Role of Transcranial Motor Evoked Potential Monitoring During Traumatic Spinal Injury Surgery: A Prospective Multicenter Study of the Monitoring Committee of the Japanese Society for Spine Surgery and Related Research.

Ushirozako Hiroki, Yoshida Go, Imagama Shiro, Machino Masaaki, Ando Muneharu, Kawabata Shigenori, Yamada Kei, Kanchiku Tsukasa, Fujiwara Yasushi, Taniguchi Shinichirou, Iwasaki Hiroshi, Shigematsu Hideki, Tadokoro Nobuaki, Takahashi Masahito, Wada Kanichiro, Yamamoto Naoya, Funaba Masahiro, Yasuda Akimasa, Hashimoto Jun, Morito Shinji, Takatani Tsunenori, Kobayashi Kazuyoshi, Nakanishi Kazuyoshi, Kurosu Kenta, Matsuyama Yukihiro

Наблюдательное исследованиеМногоцентровое исследованиеСтатьяПеревод в очереди

Аннотация

STUDY

DESIGN

A prospective multicenter observational cohort study.

OBJECTIVE

This study aimed to investigate the role of transcranial motor evoked potential (TcMEP) monitoring during traumatic spinal injury surgery, the timing of TcMEP alerts, and intervention strategies to avoid intraoperative neurological complications.

SUMMARY OF BACKGROUND DATA: Intraoperative neuromonitoring, including TcMEP monitoring, is commonly used in high-risk spinal surgery to predict intraoperative spinal cord injury; however, little information is available on its use in traumatic spinal injury surgery.

METHODS

The TcMEP monitoring data of 350 consecutive patients who underwent traumatic spinal injury surgery (mean age, 69.3 y) between 2017 and 2021 were prospectively reviewed. In this study, a TcMEP amplitude reduction ≥70% was established as a TcMEP alert. A rescue case was defined as a case with the recovery of TcMEP amplitudes after certain procedures and without postoperative neurological complications.

RESULTS

Among the 350 patients who underwent traumatic spinal injury surgery (TcMEP derivation rate 94%), TcMEP monitoring revealed seven true-positive (TP) (2.0%), three rescues (0.9%; rescue rate 30%), 31 false-positive, one false-negative, and 287 true-negative cases, resulting in 88% sensitivity, 90% specificity, 18% positive predictive value, and 99% negative predictive value. The TP rate in patients with preoperative motor deficits was 2.9%, which was higher than that in patients without preoperative motor deficits (1.1%). The most common timing of TcMEP alerts was during decompression (40%). During decompression, suspension of surgery with intravenous steroid injection was ineffective (rescue rate, 0%), and additional decompression was effective.

CONCLUSION

Given the low prevalence of neurological complications (2.3%) and the low positive predictive value (18.4%), single usage of TcMEP monitoring during traumatic spinal injury surgery is not recommended. Further efforts should be made to reduce FP alert rates through better interpretation of multimodal Intraoperative neuromonitorings and the incorporation of anesthesiology to improve the positive predictive value.

LEVEL OF EVIDENCE: 3.

Для цитирования: Ushirozako Hiroki, Yoshida Go, Imagama Shiro, Machino Masaaki, Ando Muneharu, Kawabata Shigenori, Yamada Kei, Kanchiku Tsukasa, Fujiwara Yasushi, Taniguchi Shinichirou, Iwasaki Hiroshi, Shigematsu Hideki, Tadokoro Nobuaki, Takahashi Masahito, Wada Kanichiro, Yamamoto Naoya, Funaba Masahiro, Yasuda Akimasa, Hashimoto Jun, Morito Shinji, Takatani Tsunenori, Kobayashi Kazuyoshi, Nakanishi Kazuyoshi, Kurosu Kenta, Matsuyama Yukihiro. Role of Transcranial Motor Evoked Potential Monitoring During Traumatic Spinal Injury Surgery: A Prospective Multicenter Study of the Monitoring Committee of the Japanese Society for Spine Surgery and Related Research.. Spine. 2023. DOI: 10.1097/brs.0000000000004652.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.