Diagnostic Efficacy of a Single-Lead Ambulatory 14-Day ECG Monitor in Symptomatic Children.

Côté Jean-Marc, Chetaille Philippe, Abadir Sylvia, Gosselin Louise, Simonyan David, Dallaire Frédéric

СтатьяПеревод в очереди

Аннотация

BACKGROUND

The CardioSTAT is a single-lead ambulatory electrocardiography monitor that has been validated for use in adult patients. Recording is made through 2 electrodes positioned in a lead-I configuration, and the device allows monitoring for 2, 7, or 14 days. We sought to investigate the efficacy of this device in children with paroxysmal palpitations.

METHODS

In phase I, the quality of tracings from simultaneous CardioSTAT recordings and D1-lead recordings of a standard 12-lead electrocardiography machine in 23 children were compared. Phase II was a prospective observational cohort study comparing arrhythmia detection using the CardioSTAT vs currently used devices (24-hour Holter monitor and the Cardiomemo loop recorder) in 52 children complaining of palpitations.

RESULTS

In Phase I, all but 3 rhythm strips were correctly identified. The pacing spikes on 3 strips were not adequately identified by the observers for the CardioSTAT recording. In Phase II, symptomatic episodes were reported in 42%, 73%, and 100% of subjects during monitoring with the Holter, Cardiomemo, and CardioSTATdevices, respectively. An abnormal rhythm was detected in 13%, 23%, and 35% of subjects by the Holter, Cardiomemo, and CardioSTAT monitors, respectively. The underlying rhythm during symptomatic events was determined in 90% of cases with the CardioSTAT monitor, whereas it was determined in only 19% and 29% of cases using the Holter and Cardiomemo monitors, respectively.

CONCLUSIONS

The CardioSTAT monitor provided good-quality tracings and was superior to the 24-hour Holter monitor and the Cardiomemo loop recorder in determining the presence or absence of pathologic arrhythmia in the study cohort.

Для цитирования: Côté Jean-Marc, Chetaille Philippe, Abadir Sylvia, Gosselin Louise, Simonyan David, Dallaire Frédéric. Diagnostic Efficacy of a Single-Lead Ambulatory 14-Day ECG Monitor in Symptomatic Children.. CJC open. 2021. DOI: 10.1016/j.cjco.2021.06.011.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.