DOI 10.1186/s12872-021-02298-7

Медицинская библиотека · BMC cardiovascular disorders · 2021

Development of a target product profile for a point-of-care cardiometabolic device.

Vetter Beatrice, Beran David, Boulle Philippa, Chua Arlene, de la Tour Roberto, Hattingh Lucy, Perel Pablo, Roglic Gojka, Sampath Rangarajan, Woodman Michael, Perone Sigiriya Aebischer

Consensus StatementСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

INTRODUCTION

Multi-parameter diagnostic devices can simplify cardiometabolic disease diagnosis. However, existing devices may not be suitable for use in low-resource settings, where the burden of non-communicable diseases is high. Here we describe the development of a target product profile (TPP) for a point-of-care multi-parameter device for detection of biomarkers for cardiovascular disease and metabolic disorders, including diabetes, in primary care settings in low- and middle-income countries (LMICs).

METHODS

A draft TPP developed by an expert group was reviewed through an online survey and semi-structured expert interviews to identify device characteristics requiring refinement. The draft TPP included 41 characteristics with minimal and optimal requirements; characteristics with an agreement level for either requirement of ≤ 85% in either the survey or among interviewees were further discussed by the expert group and amended as appropriate.

RESULTS

Twenty people responded to the online survey and 18 experts participated in the interviews. Twenty-two characteristics had an agreement level of ≤ 85% in either the online survey or interviews. The final TPP defines the device as intended to be used for basic diagnosis and management of cardiometabolic disorders (lipids, glucose, HbA1c, and creatinine) as minimal requirement, and offering an expanded test menu for wider cardiometabolic disease management as optimal requirement. To be suitable, the device should be intended for level 1 healthcare settings or lower, used by minimally trained healthcare workers and allow testing using self-contained cartridges or strips without the need for additional reagents. Throughput should be one sample at a time in a single or multi-analyte cartridge, or optimally enable testing of several samples and analytes in parallel with random access.

CONCLUSION

This TPP will inform developers of cardiometabolic multi-parameter devices for LMIC settings, and will support decision makers in the evaluation of existing and future devices.

Для цитирования: Vetter Beatrice, Beran David, Boulle Philippa, Chua Arlene, de la Tour Roberto, Hattingh Lucy, Perel Pablo, Roglic Gojka, Sampath Rangarajan, Woodman Michael, Perone Sigiriya Aebischer. Development of a target product profile for a point-of-care cardiometabolic device.. BMC cardiovascular disorders. 2021. DOI: 10.1186/s12872-021-02298-7.