Intravascular Ultrasound May Not Impact Graft Sizing in Endovascular Repair of Blunt Thoracic Aortic Injury.
Аннотация
OBJECTIVES
Thoracic endovascular aortic repair (TEVAR) is the preferred treatment for severe blunt thoracic aortic injuries (BTAI). Successful outcomes rely on accurate endograft sizing, but initial imaging may underestimate aortic diameters. This study examines the impact of intravascular ultrasound (IVUS) on endograft sizing and clinical outcomes in BTAI patients.
METHODS
A prospectively collected multi-institutional dataset from the Aortic Trauma Foundation was analyzed. Patients with BTAI undergoing TEVAR with IVUS were compared to patients who underwent TEVAR alone. Demographics and operative variables were compared, focusing on IVUS effects on endograft sizing by examining maximal proximal and distal aortic diameter on initial CT imaging compared to the graft diameters used during TEVAR.
RESULTS
293 patients underwent TEVAR for BTAI with IVUS utilized in 124 cases (42.3%). The average graft size in the IVUS and non-IVUS groups were similar proximally (26.91 ± 4.3 mm IVUS vs 27.77 ± 4.7 mm non-IVUS, P = 0.116) and distally (25.96 ± 4.7 mm IVUS vs 26.51 ± 4.7 mm non-IVUS). IVUS did not impact the difference between graft size and initial CT measurements proximally (4.32 ± 4.8 mm IVUS vs 4.23 ± 3.9 mm non-IVUS, P = 0.859) or distally (4.17 ± 5.9 mm IVUS vs 4.50 ± 4.3 mm non-IVUS, P = 0.606). Although delayed hemorrhagic and ischemic stroke occurred less frequently in IVUS patients (0.8% IVUS vs 7.1% non- IVUS, P = 0.024), in-hospital mortality was similar between groups (5.6% IVUS vs 7.7% non-IVUS, P = 0.581).
CONCLUSIONS
IVUS is not associated with significant changes in endograft sizing compared to sizing based on CT scan alone in BTAI patients. IVUS was not associated with differences in mortality but was associated with a decrease in delayed hemorrhagic and ischemic stroke. Routine IVUS in BTAI patients may not be necessary for accurate sizing, but there may be a relationship between IVUS and stroke.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.