Mammographic density: Comparison of visual assessment with fully automatic calculation on a multivendor dataset.
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Sacchetto Daniela, Morra Lia, Agliozzo Silvano, Bernardi Daniela, Björklund Tomas, Brancato Beniamino, Bravetti Patrizia, Carbonaro Luca A, Correale Loredana, Fantò Carmen, Favettini Elisabetta, Martincich Laura, Milanesio Luisella, Mombelloni Sara, Monetti Francesco, Morrone Doralba, Pellegrini Marco, Pesce Barbara, Petrillo Antonella, Saguatti Gianni, Stevanin Carmen, Trimboli Rubina M, Tuttobene Paola, Valentini Marvi, Marra Vincenzo, Frigerio Alfonso, Bert Alberto, Sardanelli Francesco
Аннотация
METHODS: Twenty-one radiologists assessed 613 screening/diagnostic digital mammograms from nine centers and six different vendors, using the BI-RADS a, b, c, and d density classification. The same mammograms were also evaluated by an ABDE providing the ratio between fibroglandular and total breast area on a continuous scale and, automatically, the BI-RADS score. A panel majority report (PMR) was used as reference standard. Agreement (κ) and accuracy (proportion of cases correctly classified) were calculated for binary (BI-RADS a-b versus c-d) and 4-class classification.
RESULTS: While the agreement of individual radiologists with the PMR ranged from κ = 0.483 to κ = 0.885, the ABDE correctly classified 563/613 mammograms (92 %). A substantial agreement for binary classification was found for individual reader pairs (κ = 0.620, standard deviation [SD] = 0.140), individual versus PMR (κ = 0.736, SD = 0.117), and individual versus ABDE (κ = 0.674, SD = 0.095). Agreement between ABDE and PMR was almost perfect (κ = 0.831).
CONCLUSIONS: The ABDE showed an almost perfect agreement with a 21-radiologist panel in binary BD classification on a multivendor dataset, earning a chance as a reproducible alternative to visual evaluation.
KEY POINTS: Individual BD assessment differs from PMR with κ as low as 0.483. An ABDE correctly classified 92 % of mammograms with almost perfect agreement (κ = 0.831). An ABDE can be a valid alternative to subjective BD assessment.