DOI 10.1139/apnm-2025-0482

Медицинская библиотека · Applied physiology, nutrition, and metabolism = Physiologie appliquee, nutrition et metabolisme · 2026

Reliability and validity of portable sweat sodium analyzers during exercise in the heat.

Zhao Xiujing, McKenna Zachary J, Wierick Shawn C, Perez Rosie I, Sellner Ryan, Allen Becca M, McDermott Brendon P

Клиническое исследованиеСтатьяВалидационное исследованиеПеревод в очереди

Аннотация

We aimed to evaluate reliability and validity of MX3 and LAQUAtwin-Na-11 (LANa) analyzers in measuring sweat [Na+]. Nineteen healthy, active adults (9 females, 33 ± 10 years) completed two matched, self-paced 60 min cycling or running trials in warm-humid conditions (27.3 ± 0.4 °C, 50.9 ± 1.8% RH). Sweat was collected from an absorbent patch on the left ventral forearm. Sweat was analyzed using MX3, LANa, and Medica analyzers. The accuracy of both portable analyzers was compared to Medica results. Reliability was assessed using coefficient of variation (CV) and intraclass correlation coefficient (ICC). MX3 demonstrated acceptable-to-high between-trial reliability (CV: 23%; ICC: 0.826), while LANa showed acceptable-to-moderate between-trial reliability (CV: 24%; ICC: 0.744). Validity was evaluated using CV and linear regression. Both analyzers demonstrated strong associations with the reference method (MX3: R2 = 0.867; LANa: R2 = 0.852), with good CV values (MX3: 17.5%; LANa: 13%). Overall, MX3 demonstrated higher reliability than LANa, and both analyzers provide good validity for laboratory-based sweat [Na+] assessment. Clinical registration number: NCT07490756.

Для цитирования: Zhao Xiujing, McKenna Zachary J, Wierick Shawn C, Perez Rosie I, Sellner Ryan, Allen Becca M, McDermott Brendon P. Reliability and validity of portable sweat sodium analyzers during exercise in the heat.. Applied physiology, nutrition, and metabolism = Physiologie appliquee, nutrition et metabolisme. 2026. DOI: 10.1139/apnm-2025-0482.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.