Gerstenfeld Edward P, Mansour Moussa, Whang William, Venkateswaran Ramkumar, Harding John D, Ellis Christopher R, Ellenbogen Kenneth A, Osorio Jose, DeLurgio David B, Daccarett Marcos, Mangrum Michael, McElderry Tom, Richards Elizabeth, Albrecht Elizabeth M, Schneider Christopher W, Sutton Brad S, Reddy Vivek Y

Сравнительное исследованиеСтатьяМногоцентровое исследованиеРандомизированное исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Autonomic denervation is an ancillary phenomenon during thermal ablation of atrial fibrillation (AF), that may have synergistic effects on symptomatic improvement and long-term freedom from AF. Pulsed field ablation (PFA), a nonthermal ablation modality, was noninferior to thermal ablation in treating AF; however, PFA's relative myocardial selectivity may minimize autonomic effects.

OBJECTIVES

This study sought to compare heart rate (HR) and heart rate variability (HRV) metrics as markers of autonomic function after ablation using PFA vs thermal ablation.

METHODS

ADVENT (FARAPULSE ADVENT PIVOTAL Trial PFA System vs SOC Ablation for Paroxysmal Atrial Fibrillation) was a randomized pivotal study comparing PFA (pentaspline catheter) with thermal ablation (radiofrequency [RF] or cryoballoon [CB]) for treating paroxysmal AF. Baseline HR was acquired from a pre-ablation 12-lead electrocardiogram, whereas follow-up HRs, as well as HRV (standard deviation of all normal to normal RR intervals, standard deviation of 5-minute average RR intervals) metrics, were derived from 72-hour Holter monitors at 6 and 12 months.

RESULTS

This study included 379 paroxysmal AF patients undergoing PFA (n = 194) or thermal ablation (n = 185; n = 102 RF, n = 83 CB) completing 6- and 12-month Holter monitoring. Compared with PFA, thermal patients had significantly greater increases in HR from baseline to 6 months (ΔHR; 10.1 vs 5.9 beats/min; P = 0.02) and 12 months (ΔHR; 8.8 vs 5.2 beats/min; P = 0.03). This increase in HR at 6 and 12 months was similar between CB and RF (P = 0.94 and 0.83, respectively). HRV, both standard deviation of all normal to normal RR intervals and standard deviation of 5-minute average RR intervals, were significantly lower at both 6 and 12 months after thermal ablation compared with PFA (P < 0.01).

CONCLUSIONS

PFA's effect on the autonomic nervous system was attenuated compared with thermal ablation. Whether this affects long-term freedom from AF or symptomatic bradycardia/pauses after AF ablation requires further study.

Для цитирования: Gerstenfeld Edward P, Mansour Moussa, Whang William, Venkateswaran Ramkumar, Harding John D, Ellis Christopher R, Ellenbogen Kenneth A, Osorio Jose, DeLurgio David B, Daccarett Marcos, Mangrum Michael, McElderry Tom, Richards Elizabeth, Albrecht Elizabeth M, Schneider Christopher W, Sutton Brad S, Reddy Vivek Y. Autonomic Effects of Pulsed Field vs Thermal Ablation for Treating Atrial Fibrillation: Subanalysis of ADVENT.. JACC. Clinical electrophysiology. 2024. DOI: 10.1016/j.jacep.2024.05.005.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.