DOI 10.1016/j.burns.2008.09.013

Медицинская библиотека · Burns : journal of the International Society for Burn Injuries · 2009

Validation of the Pliance X System in measuring interface pressure generated by pressure garment.

Lai Candy H Y, Li-Tsang Cecilia W P

Клиническое исследование I фазыКлиническое исследование II фазыСтатьяВалидационное исследованиеПеревод в очереди

Аннотация

INTRODUCTION

Pressure monitoring is crucial for effective pressure therapy. Precise and reliable interface pressure measurement system, however, remains unavailable in current practice.

METHODOLOGY: This is a validation study on the application of a recently developed device, the Pliance X System, for static interface pressure measurement. Sensor properties were evaluated through series of laboratory tests in the first phase of the study. Phase II was a clinical study using the Pliance X system to differentiate the loading generated on the patients' scars through additional inserts and pressure garment.

RESULTS

Results showed high test-retest and inter-rater reliability (ICC > or = 0.995) with good linearity (adjusted R-square=0.997) by measuring standard weights. The maximum deviation between the measurement generated by the sensor and the sphygmomanometer cuff was +/-1.451 mmHg. Results in the clinical trial revealed its discriminant ability in distinguishing different levels of pressure loading on patients with scars (p<0.01).

CONCLUSION

The commercially available pressure measurement system was found to be a reliable tool for measurement of low interface pressure under static condition.

Для цитирования: Lai Candy H Y, Li-Tsang Cecilia W P. Validation of the Pliance X System in measuring interface pressure generated by pressure garment.. Burns : journal of the International Society for Burn Injuries. 2009. DOI: 10.1016/j.burns.2008.09.013.