Washout times of desflurane, sevoflurane and isoflurane from the GE Healthcare Aisys® and Avance®, Carestation®, and Aestiva® anesthesia system.

Kim Tae W, Wingate Jamie R, Fernandez Allison M, Whitaker Emmett, Tham Robert Q

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND/AIM: Malignant hyperthermia susceptible patients may experience a fatal reaction to volatile anesthetic gases. This study sought to determine the washout characteristics of desflurane, sevoflurane, and isoflurane from the Aisys® , Avance® , and Aestiva® anesthesia machines. (GE Healthcare, Madison, WI, USA).

METHODS

All machines were inspected by a GE Healthcare engineer prior to testing. The machines were primed with desflurane 7 vol%, sevoflurane 2.5 vol%, and isoflurane 1.2 vol% on three separate occasions for 2 h each with each gas. The Aisys® and Avance® were tested with and without an Advanced Breathing System (ABS™ , GE Healthcare, Madison, WI, USA). The Aestiva® was tested without modification to its breathing system. Additionally, the Aisys was evaluated with desflurane 1.2 vol% and the Avance with a preflushed fresh gas line was tested with an autoclaved ABS. After priming, disposable components of the patient breathing system were replaced. The fresh gas flow was increased to 15 lpm. Gas measurements were recorded until the concentration was 4 parts per million (p.p.m).

RESULTS

The fastest median washout time was achieved by the Avance in 3 min or less without an ABS or with an autoclaved ABS. The longest median time was 35 min for the Aestiva® . Clearance of desflurane was the most time consuming for all machines.

CONCLUSION

This study demonstrates that saturated vapor pressure and priming concentration exert a greater effect on washout times than gas solubility. The Aisys utilizes an electronic vaporizer system that may expose the breathing system to retained saturated vapor. The breathing systems for all machines may hinder washout of gases.

Для цитирования: Kim Tae W, Wingate Jamie R, Fernandez Allison M, Whitaker Emmett, Tham Robert Q. Washout times of desflurane, sevoflurane and isoflurane from the GE Healthcare Aisys® and Avance®, Carestation®, and Aestiva® anesthesia system.. Paediatric anaesthesia. 2013. DOI: 10.1111/pan.12234.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.