MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s11999-012-2684-x

Медицинская библиотека · Clinical orthopaedics and related research · 2012

Adverse events after recombinant human BMP2 in nonspinal orthopaedic procedures.

Woo Emily Jane

СтатьяПеревод в очереди

Аннотация

BACKGROUND

The FDA has approved recombinant human bone morphogenetic protein 2 (rhBMP-2) for treating acute, open tibial shaft fractures. However, the nature and frequency of complications after the use of rhBMP-2 in nonspinal orthopaedic surgery have not been well characterized.

QUESTIONS/PURPOSES: To determine what types of adverse events have been reported after the use of rhBMP-2, whether they were severe enough to require additional surgery, and after what types of operations these adverse events occurred.

METHODS

Adverse events reported to the FDA's Manufacturer and User Facility Device Experience database were reviewed and summarized.

RESULTS

Through December 31, 2011, the FDA has received 62 reports of adverse events involving rhBMP-2 in nonspinal orthopaedic procedures. Surgical site infections and other wound complications, heterotopic bone, pseudarthrosis, and local inflammation were among the most commonly reported adverse events. Almost half of the reports (30 reports; 48%) stated that the patients required secondary interventions to address the reported adverse events. The majority (49 reports; 79%) described adverse events occurring after unapproved uses, such as management of tibial plateau fractures, treatment of congenital pseudarthrosis of the tibia, and humeral reconstruction.

CONCLUSIONS

Serious adverse events can occur after the use of rhBMP-2 in nonspinal orthopaedic procedures and may necessitate additional surgery. Most events in this analysis occurred after off-label uses. Postmarketing review of adverse event reports remains an important approach for identifying potential safety concerns.

Для цитирования: Woo Emily Jane. Adverse events after recombinant human BMP2 in nonspinal orthopaedic procedures.. Clinical orthopaedics and related research. 2012. DOI: 10.1007/s11999-012-2684-x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.