Case reports: fractures of threaded cups: rare complications of a well-established implant.

Hoburg Arnd T, Mehta Manav, Tohtz Stephan, Perka Carsten

Клинический случайСтатьяПеревод в очереди

Аннотация

The use of cementless threaded cups in THA is a well-established treatment. Fractures of the cups are rare complications recorded in individual cases with material defects being discussed as the primary cause. We analyzed three cases of fractured cups. Although all three cups were well fixed to existing bone, we observed deficient osseous backing dorsocranially and abrasion particles. There were no signs of femoroacetabular impingement or infection. The cups showed corrosion debris. Scanning electron microscopic investigations showed characteristics of fretting and fretting-related corrosion. We concluded the fractures occurred because of fretting combined with inadequate bony support leading to fatigue of the material and subsequent fracture.

Для цитирования: Hoburg Arnd T, Mehta Manav, Tohtz Stephan, Perka Carsten. Case reports: fractures of threaded cups: rare complications of a well-established implant.. Clinical orthopaedics and related research. 2008. DOI: 10.1007/s11999-008-0526-7.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.