Neurocritical care · 2019
Feasibility and Safety of Transnasal High Flow Air to Reduce Core Body Temperature in Febrile Neurocritical Care Patients: A Pilot Study.
СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Ziai Wendy C, Shah Dhaval, Assis Fabrizio R, Tandri Harikrishna, Geocadin Romergryko G
Аннотация
BACKGROUND: Fever is an important determinant of prognosis following acute brain injury. Current non-pharmacologic techniques to reduce fever are limited and induce a shivering response. We investigated the safety and efficacy of a novel transnasal unidirectional high flow air device in reducing core body temperature in the neurocritical care unit (NCCU) setting.
METHODS: This pilot study included seven consecutive patients in the NCCU who were febrile (> 37.5 °C) for > 24 h despite standard non-pharmacologic and first-line antipyretic agents. Medical grade high flow air was delivered transnasally using a standard continuous positive airway pressure machine with a positive pressure of 20 cmH2O for 2 h. Core esophageal and tympanic temperature were continuously monitored.
RESULTS: Mean age was 40 ± 14 yo, and 72% (5/7 patients) were men. Five patients had intracerebral or intraventricular hemorrhage, one subject had transverse myelitis, and the remaining patient had anoxic brain injury due to a cardiac arrest. After 2 h of cooling, core temperature was significantly lower than the baseline pre-cooling temperature (37.3 ± 0.5 °C vs. 38.4 ± 0.6 °C; p < 0.002). Mean transnasal airflow rate was 57.5 ± 6.5 liters per minute. Five of the seven subjects were normothermic at the end of the 2-h period. One subject with severe hyperthermia (39.7 °C) and the other with multiple interruptions to therapy due to technical reasons did not cool. The core temperature within 30 min of cessation of airflow increased and was similar to the pre-cooling baseline temperature (38.3 ± 0.4 °C vs. 38.4 ± 0.6 °C, p = NS). Rate of core cooling was 0.6 ± 0.15 °C per hour at this flow rate. No shivering response was observed. No protocol-related adverse events occurred.
CONCLUSIONS: High flow transnasal air in a unidirectional fashion lowers core body temperature in febrile patients in the NCCU setting. No adverse events were seen, and the process showed no signs of shivering or any other serious side effects during short-term exposure. This pilot study should inform further investigation.
METHODS: This pilot study included seven consecutive patients in the NCCU who were febrile (> 37.5 °C) for > 24 h despite standard non-pharmacologic and first-line antipyretic agents. Medical grade high flow air was delivered transnasally using a standard continuous positive airway pressure machine with a positive pressure of 20 cmH2O for 2 h. Core esophageal and tympanic temperature were continuously monitored.
RESULTS: Mean age was 40 ± 14 yo, and 72% (5/7 patients) were men. Five patients had intracerebral or intraventricular hemorrhage, one subject had transverse myelitis, and the remaining patient had anoxic brain injury due to a cardiac arrest. After 2 h of cooling, core temperature was significantly lower than the baseline pre-cooling temperature (37.3 ± 0.5 °C vs. 38.4 ± 0.6 °C; p < 0.002). Mean transnasal airflow rate was 57.5 ± 6.5 liters per minute. Five of the seven subjects were normothermic at the end of the 2-h period. One subject with severe hyperthermia (39.7 °C) and the other with multiple interruptions to therapy due to technical reasons did not cool. The core temperature within 30 min of cessation of airflow increased and was similar to the pre-cooling baseline temperature (38.3 ± 0.4 °C vs. 38.4 ± 0.6 °C, p = NS). Rate of core cooling was 0.6 ± 0.15 °C per hour at this flow rate. No shivering response was observed. No protocol-related adverse events occurred.
CONCLUSIONS: High flow transnasal air in a unidirectional fashion lowers core body temperature in febrile patients in the NCCU setting. No adverse events were seen, and the process showed no signs of shivering or any other serious side effects during short-term exposure. This pilot study should inform further investigation.
Предметные рубрики MeSH
AcetaminophenAdultAgedAntipyreticsBody TemperatureCerebral HemorrhageCerebral Intraventricular HemorrhageCohort StudiesContinuous Positive Airway PressureCritical CareEsophagusFeasibility StudiesFemaleFeverHumansHypoxia, BrainMaleMiddle AgedMyelitis, TransversePilot ProjectsProspective StudiesTympanic Membrane