Evaluation of Doppler USG and CSF flow wave patterns in the diagnosis of ventriculoperitoneal shunt obstructions.
Аннотация
OBJECTIVE
In the present study, our aim was to designate standards for Doppler USG images correlating with dysfunctions in each segment of the ventriculoperitoneal (VP) shunt.
MATERIALS AND
METHODS
For this mechanical study, a platform was formed with pools filled with SF on both sides. Low-, intermediate- and high-pressure shunt mechanisms were placed on this platform. For each shunt system, flow wave pattern was evaluated. Doppler USG was performed when the shunt functioned normally, the peritoneal, ventricular ends and the exit of the valve were obstructed. The results were recorded.
RESULTS
When the distal end of the peritoneal catheter was obstructed, no flow wave pattern was observed in Doppler USG. When the exit of the valve was obstructed, no flow pattern was taken from the peritoneal end, either. When the valve was pressed on the ventricular catheter, a second wave pattern with lower amplitude and a wider base was observed, in addition to the wave pattern that was observed when the pressure on the valve was removed.
CONCLUSION
Doppler USG is an effective method for the diagnosis of shunt obstruction. The evaluation of flow wave patterns using this method instead of measuring flow rates creates a different standard for each situation.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.