Safety and efficacy of an intrinsic antitachycardia pacing algorithm in patients from Japan and South Korea: results from a cardiac device registry in the Asia Pacific region.

Goya Masahiko, Park Seung-Jung, Ando Kenji, Holbrook Reece, Iskandar Rowan, Jacobsen Luke, Mozingo Joseph D, Joung Boyoung, SPARK Registry Investigators

СтатьяНаблюдательное исследованиеМногоцентровое исследованиеПеревод в очереди

Аннотация

BACKGROUND

Antitachycardia pacing (ATP) therapy, available in modern implantable cardioverter defibrillators (ICD) and cardiac resynchronization therapy defibrillators (CRT-D), aims to terminate ventricular arrhythmias without administering high energy shocks. The intrinsic ATP (iATP) algorithm automates ATP programming in real-time, tailoring therapy based on previous ATP attempts. This study evaluated the safety, efficacy, and clinical outcomes of iATP in patients from Japan and South Korea.

METHODS

This study was a prospective, observational, multi-site registry that enrolled patients from Japan and South Korea implanted with an ICD or CRT-D device with the iATP algorithm. Patients were followed for a minimum of 12 months. Outcomes included ATP termination success, appropriate shocks, acceleration, arrhythmia-related syncope, and mortality. A post hoc unanchored matching-adjusted indirect comparison (uMAIC) was performed to compare iATP with standard ATP using published literature.

RESULTS

A total of 800 patients were enrolled. The iATP success rate for terminating all episodes was 89.2% (86.2% Generalized Estimating Equation [GEE] estimated) and 82.2% for episodes in the fast VT zone (80.9% GEE estimated). Acceleration occurred in 2.0% of episodes, and arrhythmia-related syncope was observed in 0.5% of patients. The 1-year survival rate was 96.1%, with no device-related deaths or abnormal battery depletions. The uMAIC showed iATP had higher termination efficacy across all episodes (88.1% vs. 79.3%, p < 0.001), a lower probability of appropriate shocks per episode (iATP 14.7% and ATP 31.3%, p < 0.001), and fewer accelerations per episode (2.1% vs. 4.8%, p = 0.02), with similar probability of arrhythmia-related syncope per patient (0.5% vs 0.9%, p = 0.35) and mortality (12-month Kaplan Meyer survival estimate iATP 95.4%, ATP 95.3%, p = 0.43).

CONCLUSIONS

iATP exhibited a high ventricular arrhythmia termination efficacy and a favorable safety profile. Comparison of iATP to standard ATP provides initial evidence of higher termination success, lower incidence of accelerations and appropriate shocks, and similar rates of mortality and arrhythmia-related syncope.

TRIAL REGISTRATION: ClinicalTrials.gov Identifier: NCT01524276; Japan Registry of Clinical Trials Identifier: jRCT1042200049.

Для цитирования: Goya Masahiko, Park Seung-Jung, Ando Kenji, Holbrook Reece, Iskandar Rowan, Jacobsen Luke, Mozingo Joseph D, Joung Boyoung, SPARK Registry Investigators. Safety and efficacy of an intrinsic antitachycardia pacing algorithm in patients from Japan and South Korea: results from a cardiac device registry in the Asia Pacific region.. Journal of medical economics. 2025. DOI: 10.1080/13696998.2025.2543213.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.