Insertion length and resistance during advancing of epidural catheter.

Kundra Pankaj, Viswanath Senthil Kumar, Meena Dharam S, Badhe Ashok

СтатьяРандомизированное исследованиеПеревод в очереди

Аннотация

PURPOSE: The migration of an epidural catheter into the intravascular and subarachnoid spaces sometimes occurs. This study was designed to investigate where the resistance was felt during the advancing of the catheter into the epidural space and whether the length of catheter advanced in the epidural space affected the incidence of catheter migration.

METHODS

One hundred and twenty women, American Society of Anesthesiologists (ASA) 1 or 2, scheduled to undergo lower abdominal surgeries with epidural anesthesia were randomly assigned to two groups according to the length of the epidural catheter advanced; 4 cm (n = 60) or 8 cm (n = 60). The length where resistance to advancing the catheter was perceived was recorded in all patients, and the incidence of aspiration of blood or cerebrospinal fluid (CSF) was obtained. Further, the catheters removed 48 h after surgery were scrutinized for their bending sites.

RESULTS

Resistance was felt in 83 (69.2%) patients and the mean length in the epidural space at which resistance was found was 2.5 +/- 1.2 cm. Blood was aspirated in 9 (7.5%) patients when resistance to advancing the catheter was overcome, but CSF was aspirated in no patient. A distal bend was observed 2.4 +/- 1.3 cm from the tip of the catheter, and the sites of bending were correlated with the length where resistance was encountered. An additional proximal bend was observed in 35 (58%) patients in the 8-cm group, and in 2 patients (3%) in the 4-cm group (P < 0.001), probably due to coiling of the catheter.

CONCLUSION

At approximately 2.5 cm in the epidural space, advancing an epidural catheter causes resistance. Further advancing past this point may cause migration of the catheter into the vessels, or the coiling of the catheter.

Для цитирования: Kundra Pankaj, Viswanath Senthil Kumar, Meena Dharam S, Badhe Ashok. Insertion length and resistance during advancing of epidural catheter.. Journal of anesthesia. 2009. DOI: 10.1007/s00540-009-0818-4.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.