DOI 10.1007/s00392-023-02368-0

Медицинская библиотека · Clinical research in cardiology : official journal of the German Cardiac Society · 2024

The effect of a smaller spacer in the PASCAL Ace on residual mitral valve orifice area.

Paukovitsch Michael, Felbel Dominik, Tadic Marijana, Keßler Mirjam, Scheffler Jinny, Gröger Matthias, Markovic Sinisa, Rottbauer Wolfgang, Schneider Leonhard Moritz

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Mitral transcatheter edge-to-edge repair (M-TEER) is an established treatment for functional mitral regurgitation (FMR) associated with a risk of creating iatrogenic stenosis.

OBJECTIVES

To investigate the impact of the P10 and its larger spacer compared to the narrower Ace and its smaller spacer on reduction of mitral valve orifice area (MVOA) during M-TEER.

METHODS

Consecutive patients undergoing M-TEER for treatment of severe FMR were screened retrospectively. Patients with a single PASCAL device implantation within the central segments of the MV leaflets, non-complex anatomy, and baseline MVOA ≥ 3.5cm2 were selected. Intraprocedural transesophageal echocardiography was used to compare MVOA reduction with 3D multiplanar reconstruction and direct planimetry. Device selection did not follow a prespecified MVOA threshold.

RESULTS

Seventy-two patients (81.0 years, IQR {74.3-85.0}) were included. In 32 patients, the P10 was implanted (44.4%). MR severity (p = 0.66), MR reduction (p = 0.73), and body surface area (p = 0.56) were comparable. Baseline MVOA tended to be smaller in P10 patients with the larger spacer (5.0 ± 1.1 vs. 5.4 ± 1.3cm2, p = 0.18), however, residual MVOA was larger in these patients (2.7 ± 0.7 vs. 2.3 ± 0.6cm2, p = 0.03). Accordingly, relative MVOA reduction was significantly less in P10 patients (- 45.9 ± 7.6 vs. - 56.3 ± 7.0%, p < 0.01). Indirect annuloplasty was more pronounced in Ace patients whereas mean transmitral gradients were similar.

CONCLUSION

In FMR patients with non-complex anatomy, the larger spacer of the P10 maintains greater MVOA with similar MR reduction. Hence, the use of the PASCAL Ace device in patients with small MVOAs might correlate with a risk of both clinically relevant orifice reduction and even iatrogenic stenosis.

Для цитирования: Paukovitsch Michael, Felbel Dominik, Tadic Marijana, Keßler Mirjam, Scheffler Jinny, Gröger Matthias, Markovic Sinisa, Rottbauer Wolfgang, Schneider Leonhard Moritz. The effect of a smaller spacer in the PASCAL Ace on residual mitral valve orifice area.. Clinical research in cardiology : official journal of the German Cardiac Society. 2024. DOI: 10.1007/s00392-023-02368-0.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.