Result of the left bundle branch area pacing data collection registry, an international multicenter study of conduction system pacing with a Tendril STS 2088 stylet-driven lead.

Vijayaraman Pugazhendhi, Foo David, Lim Tien Siang Eric, Diamantakos Evangelos, Verma Sumit, Hourdain Jerome, Pandurangi Ulhas, Bhargava Kartikeya, Zanon Francesco, Stone James, Lee Kwangdeok, Nair Devi

СтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

INTRODUCTION

Left bundle branch area (LBBA) pacing (LBBAP) has been proposed as an alternative therapy option in patients indicated for cardiac pacing to treat bradycardia or heart failure. The aim of the study was to evaluate the safety and effectiveness of LBBAP in patients implanted with a Tendril 2088 stylet-driven lead.

METHODS

The international retrospective data collection registry included 11 sites from 5 countries globally. Patients with attempted implants of the Tendril lead in the LBBA were followed for at least 6 months post the implant attempt. The primary safety and efficacy endpoints were freedom from LBBAP lead-related serious adverse events and the composite of LBBA capture threshold of ≤2.0 V and R-wave amplitudes ≥5 mV (or ≥value at implant), respectively.

RESULTS

Of 221 patients with attempted implants of the Tendril 2088 lead in the LBBA, 91.4% (202/221) had successful implants for LBBAP. Regardless of the LBBAP implant success, all patients were followed for at least 6 months (8.7 ± 7.3 months). Baseline characteristics: 44% female, 84% ≥65 years old, 34% coronary artery disease, and 86% of primary indications for pacemaker implant. Both primary safety and effectiveness endpoints were met (freedom from LBBAP lead-related serious adverse device effects of 99.5% and electrical performance composite success rate of 93%). The capture thresholds in LBBAP at implant and 6 months were 0.8 ± 0.3 V@0.4 ± 0.1 ms and 0.8 ± 0.3 V@0.4 ± 0.1 ms. The rate of patients with capture threshold rise ≥1 V was 1.5% through 6 months. The R-wave amplitudes in LBBAP at implant and 6 months were 9.3 ± 3.2 mV and 10.6 ± 3.0 mV.

CONCLUSIONS

This large multicenter study demonstrates that the stylet-driven Tendril™ STS 2088 lead is safe and effective for LBBAP with high success and low complication rates.

Для цитирования: Vijayaraman Pugazhendhi, Foo David, Lim Tien Siang Eric, Diamantakos Evangelos, Verma Sumit, Hourdain Jerome, Pandurangi Ulhas, Bhargava Kartikeya, Zanon Francesco, Stone James, Lee Kwangdeok, Nair Devi. Result of the left bundle branch area pacing data collection registry, an international multicenter study of conduction system pacing with a Tendril STS 2088 stylet-driven lead.. Journal of cardiovascular electrophysiology. 2024. DOI: 10.1111/jce.16274.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.