Effect of Radial Head Implant Shape on Radiocapitellar Joint Congruency.

Lalone Emily A, Shannon Hannah L, Deluce Simon R, Giles Joshua W, King Graham J W, Johnson James A

СтатьяПеревод в очереди

Аннотация

PURPOSE: Radial head arthroplasty is indicated in displaced fractures in which comminution precludes successful internal fixation. Many types of radial head implants have been developed varying in material, methods of fixation, and degrees of modularity and geometry. The purpose of this study was to investigate the effect of radial head implant shape on radiocapitellar joint congruency.

METHODS

Joint congruency was quantified in 7 cadaveric specimens employing a registration and inter-surface distance algorithm and 3-dimensional models obtained using computed tomography. Forearm rotation was simulated after computer-guided implantation of an axisymmetric radial head, a population-based quasi-anatomic radial head implant, and a reverse-engineered anatomic radial head implant. Inter-surface distances were measured to investigate the relative position of the radial head implant and displayed on 3-dimensional color-contour maps. Surface area was measured for inter-surface distances (1.5 mm) and compared for each radial head geometry.

RESULTS

There were no statistical differences in the contact surface area between radial head implants during active or passive forearm rotation. The joint was more congruent (larger contact surface area) during active forearm rotation compared with passive forearm rotation.

CONCLUSIONS

This study investigated the effect of implant geometry on the radiocapitellar joint contact mechanics by examining a commercially available radial head system (axisymmetric), a quasi-anatomic design, and an anatomic reverse-engineered radial head implant. We found no statistical differences in radiocapitellar joint contact mechanics as measured by 3-dimensional joint congruency in cadaveric specimens undergoing continuous simulated forearm rotation.

CLINICAL RELEVANCE: The importance of choosing an implant that matches the general size of the native radial head is recognized, but the degree to which it is necessary to create an implant that replicates the native anatomy to restore elbow stability and prevent cartilage degenerative changes remains unclear. This study concluded that the geometry of the implant did not have a statistically significant effect on joint contact mechanics; therefore, future work is needed to examine additional factors related to implant design, such as material choice and implant positioning to investigate their influence on joint contact mechanics.

Для цитирования: Lalone Emily A, Shannon Hannah L, Deluce Simon R, Giles Joshua W, King Graham J W, Johnson James A. Effect of Radial Head Implant Shape on Radiocapitellar Joint Congruency.. The Journal of hand surgery. 2017. DOI: 10.1016/j.jhsa.2017.03.009.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.