DOI 10.1007/s10557-024-07584-y
Enhanced Reendothelialization and Thrombosis Prevention with a New Drug-Eluting Stent.
Аннотация
PURPOSE: The objective of the study is to test the efficacy of cyclopentenyl cytosine (CPEC)-coated stents on blocking artery stenosis, promoting reendothelialization, and reducing thrombosis.
METHODS
Scanning electron microscopy was employed to observe the morphological characteristics of stents coated with a mixture of CPEC and poly(lactic-co-glycolic acid) (PLGA) copolymer. PLGA has been used in various Food and Drug Administration (FDA)-approved therapeutic devices. In vitro release of CPEC was tested to measure the dynamic drug elution. Comparison between CPEC- and everolimus-coated stents on neointimal formation and thrombosis formation was conducted after being implanted into the human internal mammary artery and grafted to the mouse aorta.
RESULTS
Optimization in stent coating resulted in uniform and consistent coating with minimal variation. In vitro drug release tests demonstrated a gradual and progressive discharge of CPEC. CPEC- or everolimus-coated stents caused much less stenosis than bare-metal stents. However, CPEC stent-implanted arteries exhibited enhanced reendothelialization compared to everolimus stents. Mechanistically, CPEC-coated stents reduced the proliferation of vascular smooth muscle cells while simultaneously promoting reendothelialization. More significantly, unlike everolimus-coated stents, CPEC-coated stents showed a significant reduction in thrombosis formation even in the absence of ongoing anticoagulant treatment.
CONCLUSIONS
The study establishes CPEC-coated stent as a promising new device for cardiovascular interventions. By enhancing reendothelialization and preventing thrombosis, CPEC offers advantages over conventional approaches, including the elimination of the need for anti-clogging drugs, which pave the way for improved therapeutic outcomes and management of atherosclerosis-related medical procedures.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.