The Journal of surgical research · 2020
Measuring Forces in Suture Techniques for Wound Closure.
Сравнительное исследованиеСтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Marsidi Nick, Vermeulen Sofieke A M, Horeman Tim, Genders Roel E
Аннотация
BACKGROUND: The use of sutures remains the first choice for wound closure. However, incorrect use of a suture technique can lead to impaired healing. Many techniques are described for high-tension wounds, but not much is known about their mechanical properties. Complications of excessive tension include dehiscence, infection, and ischemic necrosis and could be prevented. This study aimed to compare forces in five techniques (single, horizontal mattress, vertical mattress, pulley, and modified pulley suture) in a standardized wound tension model.
MATERIALS AND METHODS: A standardized neoprene wound model was developed on the ForceTRAP system (MediShield B.V., Delft, The Netherlands) to mimic a 5 Newton (N) wound. Five different suture techniques were each repeated 10 times by a student, resident dermatology, and dermsurgeon. The pulling force of the suture's first throw was measured with the Hook-in-Force sensor (Technical University Delft, The Netherlands). Changes in wound tension were measured by the ForceTRAP system. The ForceTRAP is a platform measuring forces from 0 to 20 N in three dimensions with an accuracy of 0.1 N. The Hook-in-Force is a force sensor measuring 0-15 N with an accuracy of 0.5 N. Maximum and mean forces were calculated for each suture technique and operator.
RESULTS: Mean maximum pulling force: 5.69 N (standard deviation [SD], 0.88) single, 7.25 N (SD, 1.33) vertical mattress, 8.11 N (SD, 1.00) horizontal mattress, 3.46 N (SD, 0.61) pulley, and 4.52 N (SD, 0.67) modified pulley suture. The mean force increase on the skin (substitute) ranged between 0.80 N (pulley) and 0.96 N (vertical mattress).
CONCLUSIONS: The pulley suture requires less pulling force compared with other techniques. The mechanical properties of sutures should be taken in consideration when choosing a technique to close wounds.
MATERIALS AND METHODS: A standardized neoprene wound model was developed on the ForceTRAP system (MediShield B.V., Delft, The Netherlands) to mimic a 5 Newton (N) wound. Five different suture techniques were each repeated 10 times by a student, resident dermatology, and dermsurgeon. The pulling force of the suture's first throw was measured with the Hook-in-Force sensor (Technical University Delft, The Netherlands). Changes in wound tension were measured by the ForceTRAP system. The ForceTRAP is a platform measuring forces from 0 to 20 N in three dimensions with an accuracy of 0.1 N. The Hook-in-Force is a force sensor measuring 0-15 N with an accuracy of 0.5 N. Maximum and mean forces were calculated for each suture technique and operator.
RESULTS: Mean maximum pulling force: 5.69 N (standard deviation [SD], 0.88) single, 7.25 N (SD, 1.33) vertical mattress, 8.11 N (SD, 1.00) horizontal mattress, 3.46 N (SD, 0.61) pulley, and 4.52 N (SD, 0.67) modified pulley suture. The mean force increase on the skin (substitute) ranged between 0.80 N (pulley) and 0.96 N (vertical mattress).
CONCLUSIONS: The pulley suture requires less pulling force compared with other techniques. The mechanical properties of sutures should be taken in consideration when choosing a technique to close wounds.