MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1080/15298667991430451

Медицинская библиотека · American Industrial Hygiene Association journal · 1979

A sampling and analytical method for vinylidene chloride in air.

Foerst D

СтатьяПеревод в очереди

Аннотация

A sampling and analytical method for the determination of vinylidene chloride in air is presented. Vinylidene chloride can be quantitatively trapped on charcoal over a wide range of concentrations. The collected vapors are desorbed with carbon disulfide and analyzed by gas chromatography utilizing a flame ionization detector. The analytical column is packed with Durapak OPN. The lower limit for the method is approximately 7 micrograms of vinylidene chloride per sampling tube. The breakthrough volume was found to be dependent upon the concentration sampled and the relative humidity of the air sampled. The overall precision of the analytical method is 5% (relative standard deviation).

Для цитирования: Foerst D. A sampling and analytical method for vinylidene chloride in air.. American Industrial Hygiene Association journal. 1979. DOI: 10.1080/15298667991430451.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.