DOI 10.1007/s11325-019-01904-5
Detecting central sleep apnea in adult patients using WatchPAT-a multicenter validation study.
Аннотация
STUDY
OBJECTIVES
To assess the accuracy of WatchPAT (WP-Itamar-Medical, Caesarea, Israel) enhanced with a novel systolic upstroke analysis coupled with respiratory movement analysis derived from a dedicated snoring and body position (SBP) sensor, to enable automated algorithmic differentiation between central sleep apnea (CSA) and obstructive sleep apnea (OSA) compared with simultaneous in-lab sleep studies with polysomnography (PSG).
METHODS
Eighty-four patients with suspected sleep-disordered breathing (SDB) underwent simultaneous WP and PSG studies in 11 sleep centers. PSG scoring was blinded to the automatically analyzed WP data.
RESULTS
Overall WP apnea-hypopnea index (AHI; mean ± SD) was 25.2 ± 21.3 (range 0.2-101) versus PSG AHI 24.4 ± 21.2 (range 0-110) (p = 0.514), and correlation was 0.87 (p < 0.001). Using a threshold of AHI ≥ 15, the sensitivity and specificity of WP versus PSG for diagnosing sleep apnea were 85% and 70% respectively and agreement was 79% (kappa = 0.867). WP central AHI (AHIc) was 4.2 ± 7.7 (range 0-38) versus PSG AHIc 5.9 ± 11.8 (range 0-63) (p = 0.034), while correlation was 0.90 (p < 0.001). Using a threshold of AHI ≥ 15, the sensitivity and specificity of WP versus PSG for diagnosing CSA were 67% and 100% respectively with agreement of 95% (kappa = 0.774), and receiver operator characteristic (ROC) area under the curve of 0.866, (p < 0.01). Using a threshold of AHI ≥ 10 showed comparable overall sleep apnea and CSA diagnostic accuracies.
CONCLUSIONS
These findings show that WP can accurately detect overall AHI and effectively differentiate between CSA and OSA.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.