Ultra-low current electrospray ionization of chloroform solution for the analysis of perfluorinated sulfonic acids.

СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Wang Taoqing, Li Huishan, Allen Nicholas, Ferraro Ian, Li Anyin

Аннотация

RATIONALE: Femtoamp and picoamp electrospray ionization (ESI) characteristics of a nonpolar solvent were explored. The direct ESI mass spectrometry analysis of chloroform extract solution enabled rapid analysis of perfluorinated sulfonic acid analytes in drinking water.

METHODS: Neat chloroform solvent and extracts were directly used in a typical wire-in ESI setup using micrometer emitter tips. Ionization currents were measured with femtoamp sensitivity while ramping the spray voltage from 0 to -5000 V. Methanol was used as a comparison to illustrate the characteristics of electrospraying chloroform. The effects of spray voltage and inlet temperature were studied. A liquid-liquid extraction workflow was developed to analyze perfluorooctanoate sulfonate (PFOS) in drinking water using an ion-trap mass spectrometer.

RESULTS: The ionization onset of chloroform solution was 41 ± 17 fA at 300 V. The ionization current gradually increased with voltage while remaining below 100 pA when using voltages up to -5000 V. The ion signal of PFOS was significantly enhanced to improve the limit of detection (LoD) to 25 ppt in chloroform. Coupled with a liquid-liquid extraction workflow, LoD of 0.38-5.1 ppt and a quantitation range of 5-400 ppt were achieved for perfluorinated sulfonic compounds in 1-ml water samples.

CONCLUSIONS: Femtoamp and picoamp modes expand the solvent compatibility range of ESI and can enable quantitative analysis in parts per trillion (ppt) concentrations.