Ultra-low current electrospray ionization of chloroform solution for the analysis of perfluorinated sulfonic acids.

Wang Taoqing, Li Huishan, Allen Nicholas, Ferraro Ian, Li Anyin

СтатьяПеревод в очереди

Аннотация

RATIONALE: Femtoamp and picoamp electrospray ionization (ESI) characteristics of a nonpolar solvent were explored. The direct ESI mass spectrometry analysis of chloroform extract solution enabled rapid analysis of perfluorinated sulfonic acid analytes in drinking water.

METHODS

Neat chloroform solvent and extracts were directly used in a typical wire-in ESI setup using micrometer emitter tips. Ionization currents were measured with femtoamp sensitivity while ramping the spray voltage from 0 to -5000 V. Methanol was used as a comparison to illustrate the characteristics of electrospraying chloroform. The effects of spray voltage and inlet temperature were studied. A liquid-liquid extraction workflow was developed to analyze perfluorooctanoate sulfonate (PFOS) in drinking water using an ion-trap mass spectrometer.

RESULTS

The ionization onset of chloroform solution was 41 ± 17 fA at 300 V. The ionization current gradually increased with voltage while remaining below 100 pA when using voltages up to -5000 V. The ion signal of PFOS was significantly enhanced to improve the limit of detection (LoD) to 25 ppt in chloroform. Coupled with a liquid-liquid extraction workflow, LoD of 0.38-5.1 ppt and a quantitation range of 5-400 ppt were achieved for perfluorinated sulfonic compounds in 1-ml water samples.

CONCLUSIONS

Femtoamp and picoamp modes expand the solvent compatibility range of ESI and can enable quantitative analysis in parts per trillion (ppt) concentrations.

Для цитирования: Wang Taoqing, Li Huishan, Allen Nicholas, Ferraro Ian, Li Anyin. Ultra-low current electrospray ionization of chloroform solution for the analysis of perfluorinated sulfonic acids.. Rapid communications in mass spectrometry : RCM. 2023. DOI: 10.1002/rcm.9501.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.