DOI 10.1542/peds.2013-1713

Медицинская библиотека · Pediatrics · 2013

Reliability of transcutaneous bilirubin devices in preterm infants: a systematic review.

Nagar Gaurav, Vandermeer Ben, Campbell Sandra, Kumar Manoj

СтатьяМетаанализПоддержка исследования: вне правительства СШАСистематический обзорПеревод в очереди

Аннотация

BACKGROUND AND

OBJECTIVE

Transcutaneous bilirubin (TcB) devices are widely used for the estimation of serum bilirubin levels in term and near-term infants. Our objective was to review the diagnostic accuracy of TcB devices in preterm infants.

METHODS

Medline, Embase, Cochrane library, Cumulative Index to Nursing and Allied Health Literature, and Scopus were searched (from database inception date until December 2012). Additional citations were identified by using the bibliographies of selected articles and from conference proceedings. The studies were included if they compared TcB with total serum bilirubin in preterm infants before phototherapy and presented data as correlation coefficients or as Bland-Altman difference plots. Data were extracted by 1 reviewer and checked for accuracy by the second reviewer. An assessment tool (quality assessment of diagnostic accuracy studies) was used for risk of bias assessments.

RESULTS

Twenty-two studies met the inclusion criteria; 21 studies reported results as correlation coefficients, with pooled estimates of r = 0.83 for each site of measurement. Pooled estimates in infants <32 weeks' gestation were similar to the overall preterm population (r = 0.89 [95% confidence interval: 0.82-0.93]). For the 2 commonly used TcB devices (ie, JM103 and BiliCheck), the results were comparable at the forehead site, although the JM103 device exhibited better correlation at the sternum. Analysis of the Bland-Altman plots (13 studies) revealed negligible bias in measurement at the forehead or sternum site by using either the JM-103 or BiliCheck device; however, the JM-103 device exhibited better precision than the BiliCheck (SD for TcB - total serum bilirubin differences: 24.3 and 31.98 µmol/L, respectively).

CONCLUSIONS

The TcB devices reliably estimated bilirubin levels in preterm infants and could be used in clinical practice to reduce blood sampling.

Для цитирования: Nagar Gaurav, Vandermeer Ben, Campbell Sandra, Kumar Manoj. Reliability of transcutaneous bilirubin devices in preterm infants: a systematic review.. Pediatrics. 2013. DOI: 10.1542/peds.2013-1713.