In vitro investigation of axial mechanical support devices implanted in the novel convergent cavopulmonary connection Fontan.

Cleveland Vincent, Contento Jacqueline, Mass Paige, Hardikar Priyanka, Wu Qiyuan, Liu Xiaolong, Aslan Seda, Loke Yue-Hin, Krieger Axel, Lunos Scott, Olivieri Laura, Sinha Pranava

СтатьяПеревод в очереди

Аннотация

OBJECTIVES

The 2 opposing inflows and 2 outflows in a total cavopulmonary connection make mechanical circulatory support (MCS) extremely challenging. We have previously reported a novel convergent cavopulmonary connection (CCPC) Fontan design that improves baseline characteristics and provides a single inflow and outflow, thus simplifying MCS. This study aims to assess the feasibility of MCS of this novel configuration using axial flow pumps in an in vitro benchtop model.

METHODS

Three-dimensional segmentations of 12 single-ventricle patients (body surface area 0.5-1.75 m2) were generated from cardiovascular magnetic resonance images. The CCPC models were designed by connecting the inferior vena cava and superior vena cava to a shared conduit ascending to the pulmonary arteries, optimized in silico. The 12 total cavopulmonary connection and their corresponding CCPC models underwent in vitro benchtop characterization. Two MCS devices were used, the Impella RP® and the PediPump.

RESULTS

MCS successfully and symmetrically reduced the pressure in both vena cavae by >20 mmHg. The devices improved the hepatic flow distribution balance of all CCPC models (Impella RP®P = 0.045, PediPump P = 0.055).

CONCLUSIONS

The CCPC Fontan design provides a feasible MCS solution for a failing Fontan by balancing hepatic flow distribution and symmetrically decompressing the central venous pressure. Cardiac index may also improve with MCS. Additional studies are needed to evaluate this concept for managing Fontan failure.

Для цитирования: Cleveland Vincent, Contento Jacqueline, Mass Paige, Hardikar Priyanka, Wu Qiyuan, Liu Xiaolong, Aslan Seda, Loke Yue-Hin, Krieger Axel, Lunos Scott, Olivieri Laura, Sinha Pranava. In vitro investigation of axial mechanical support devices implanted in the novel convergent cavopulmonary connection Fontan.. European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery. 2024. DOI: 10.1093/ejcts/ezad413.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.