The Journal of hand surgery · 2017
Fixation of Regan-Morrey Type II Coronoid Fractures: A Comparison of Screws and Suture Lasso Technique for Resistance to Displacement.
Сравнительное исследованиеСтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Iannuzzi Nicholas P, Paez Adrian G, Parks Brent G, Murphy Michael S
Аннотация
PURPOSE: The aim of this study was to compare the load to failure and stiffness achieved in coronoid fractures treated with a posterior-to-anterior screw versus a suture lasso technique.
METHODS: We performed a biomechanical study using 10 pairs of fresh-frozen cadaveric elbows. A transverse osteotomy at the midpoint of the coronoid height was created to simulate a Regan-Morrey type II coronoid fracture. The specimens were randomized to screw fixation or suture lasso fixation. The load to failure and stiffness were then measured using a materials testing machine.
RESULTS: Screw fixation provided greater strength and stiffness than suture lasso fixation. Mean load to failure was 405 N in the screw fixation group compared with a load to failure of 207 N for suture fixation. Screw fixation resulted in a mean stiffness of 284 kPa/mm compared with 119 kPa/mm after suture fixation.
CONCLUSIONS: Screw fixation was biomechanically superior to fixation using a suture lasso technique. For coronoid fractures in which screw or suture fixation is feasible, screw fixation may provide greater resistance to displacement of the coronoid compared with a suture lasso technique.
CLINICAL RELEVANCE: Clinical studies have reported a higher rate of failure after screw fixation compared with suture lasso fixation; however, this study demonstrated a greater stiffness and load to failure after screw fixation of type II coronoid fractures. Screw fixation may provide a stronger fixation construct for fractures of adequate size to support a screw. Further studies may be warranted to assess the importance of securing the anterior capsule to the coronoid tip when using a suture lasso construct because this may affect the stability of the elbow after fixation.
METHODS: We performed a biomechanical study using 10 pairs of fresh-frozen cadaveric elbows. A transverse osteotomy at the midpoint of the coronoid height was created to simulate a Regan-Morrey type II coronoid fracture. The specimens were randomized to screw fixation or suture lasso fixation. The load to failure and stiffness were then measured using a materials testing machine.
RESULTS: Screw fixation provided greater strength and stiffness than suture lasso fixation. Mean load to failure was 405 N in the screw fixation group compared with a load to failure of 207 N for suture fixation. Screw fixation resulted in a mean stiffness of 284 kPa/mm compared with 119 kPa/mm after suture fixation.
CONCLUSIONS: Screw fixation was biomechanically superior to fixation using a suture lasso technique. For coronoid fractures in which screw or suture fixation is feasible, screw fixation may provide greater resistance to displacement of the coronoid compared with a suture lasso technique.
CLINICAL RELEVANCE: Clinical studies have reported a higher rate of failure after screw fixation compared with suture lasso fixation; however, this study demonstrated a greater stiffness and load to failure after screw fixation of type II coronoid fractures. Screw fixation may provide a stronger fixation construct for fractures of adequate size to support a screw. Further studies may be warranted to assess the importance of securing the anterior capsule to the coronoid tip when using a suture lasso construct because this may affect the stability of the elbow after fixation.