Intraoperative pulmonary barotrauma during ophthalmologic surgery. Case report.
Клинический случайСтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Zugliani Affonso Henrique, Claro Flávia, Mega Ana Cláudia C, Rodrigues Marcelo F, Ancelmé Gilblainer
Аннотация
BACKGROUND AND OBJECTIVES: Nowadays, severe anesthetic complications caused by the improper use of mechanical ventilators are rare. However, technical details even in recent models can be a trap for the anesthesiologist and threaten patient safety. The objective of this report was to demonstrate the importance of a careful analysis of the device to be used, as well as to detect and treat intraoperative tension pneumothorax.
CASE REPORT: A 16-year old female patient, physical status ASA I, underwent corneal conjunctival covering under general anesthesia. Anesthesia was maintained with isoflurane and controlled mechanical ventilation. No abnormalities were observed during anesthesia. At the final phase of the surgery, after mobilizing the anesthesia device to start the awakening process, the patient developed hypoxia, hypertension and ventilatory difficulties. After removal of the sterile drapes from the surgical field, subcutaneous emphysema was evident in the face, neck and upper limb. The tracheal cannula, which contained blood, was changed. A chest Xray confirmed the diagnosis of pneumothorax that was immediately drained. Inspection of the equipment revealed the presence of a kink in the tubing connecting the inferior portion of the canister to the equipment itself caused by mobilization of the articulated arm, blocking the normal flow of gases and leading to pulmonary barotrauma.
CONCLUSIONS: The development of tension pneumothorax during general anesthesia with positive pressure ventilation should always be considered. Several factors can contribute to the development of this condition, which should be considered when they are present during surgeries. The anesthesia equipment should be examined carefully to detect potential causes of anesthetic complications.
CASE REPORT: A 16-year old female patient, physical status ASA I, underwent corneal conjunctival covering under general anesthesia. Anesthesia was maintained with isoflurane and controlled mechanical ventilation. No abnormalities were observed during anesthesia. At the final phase of the surgery, after mobilizing the anesthesia device to start the awakening process, the patient developed hypoxia, hypertension and ventilatory difficulties. After removal of the sterile drapes from the surgical field, subcutaneous emphysema was evident in the face, neck and upper limb. The tracheal cannula, which contained blood, was changed. A chest Xray confirmed the diagnosis of pneumothorax that was immediately drained. Inspection of the equipment revealed the presence of a kink in the tubing connecting the inferior portion of the canister to the equipment itself caused by mobilization of the articulated arm, blocking the normal flow of gases and leading to pulmonary barotrauma.
CONCLUSIONS: The development of tension pneumothorax during general anesthesia with positive pressure ventilation should always be considered. Several factors can contribute to the development of this condition, which should be considered when they are present during surgeries. The anesthesia equipment should be examined carefully to detect potential causes of anesthetic complications.