Mostafa Mohamed A, Bodziock George, Cotten Lindsey, Schaich Christopher L, Seiler Amber, Dillon John, Brock Jonathan, Hansen Ross, Kozak Patrick, Simmons Tony, Bradford Natalie, Allred James, Whalen Patrick, Bhave Prashant D
Аннотация
BACKGROUND: The rapid evolution of cardiac implantable electronic devices (CIEDs) has increased remote transmission data, leading to excessive non-actionable alerts (NAA) and alert fatigue.
OBJECTIVE: Optimize alert parameters to minimize NAA and evaluate the impact on clinical outcomes.
METHODS: We included 536 participants (mean age 75 (15) years, 60.4% male, 83.4% white) with CIEDs. In 413 patients, CIEDs were reprogrammed to censor alerts as follows: atrial fibrillation (AF) episodes < 5.5 h, persistent AF > 1 month with prior alerts, AF < 24 h on anticoagulation or with prior appendage occlusion, and non-sustained ventricular tachycardia (NSVT) in defibrillator platforms. NAAs were tracked 90-days pre- and post-reprogramming. Incident ischemic stroke and sudden cardiac death (SCD) were assessed over a median 1.8-year follow-up. Logistic regression models examined associations between reprogramming and outcomes.
RESULTS: Reprogramming was implemented for AF alerts (69.5%, n = 287) and NSVT alerts (30.5%, n = 126). After reprogramming, NAAs significantly decreased from 6.68 (SD = 10.02) to 2.27 (SD = 4.58), p < 0.001. During follow-up, ischemic stroke rates in AF patients were similar between reprogrammed (5.2%, n = 15) and control groups (5.4%, n = 5). In those with NSVT alerts, SCD incidence was lower in reprogrammed (2.3%, n = 3) versus controls (9.3%, n = 3). In logistic regression models adjusted for demographics, CHA₂DS₂VASC score, anticoagulation status, and prior stroke history, there was no statistically significant difference in stroke risk between groups (OR 0.82 [0.27-2.51]).
CONCLUSIONS: Guideline-based alert parameters in CIED patients significantly reduced NAA burden with no increasing in adverse outcomes in patients with device-detected AF or NSVT alerts. This approach may reduce noise and safely improve efficiency.
OBJECTIVE: Optimize alert parameters to minimize NAA and evaluate the impact on clinical outcomes.
METHODS: We included 536 participants (mean age 75 (15) years, 60.4% male, 83.4% white) with CIEDs. In 413 patients, CIEDs were reprogrammed to censor alerts as follows: atrial fibrillation (AF) episodes < 5.5 h, persistent AF > 1 month with prior alerts, AF < 24 h on anticoagulation or with prior appendage occlusion, and non-sustained ventricular tachycardia (NSVT) in defibrillator platforms. NAAs were tracked 90-days pre- and post-reprogramming. Incident ischemic stroke and sudden cardiac death (SCD) were assessed over a median 1.8-year follow-up. Logistic regression models examined associations between reprogramming and outcomes.
RESULTS: Reprogramming was implemented for AF alerts (69.5%, n = 287) and NSVT alerts (30.5%, n = 126). After reprogramming, NAAs significantly decreased from 6.68 (SD = 10.02) to 2.27 (SD = 4.58), p < 0.001. During follow-up, ischemic stroke rates in AF patients were similar between reprogrammed (5.2%, n = 15) and control groups (5.4%, n = 5). In those with NSVT alerts, SCD incidence was lower in reprogrammed (2.3%, n = 3) versus controls (9.3%, n = 3). In logistic regression models adjusted for demographics, CHA₂DS₂VASC score, anticoagulation status, and prior stroke history, there was no statistically significant difference in stroke risk between groups (OR 0.82 [0.27-2.51]).
CONCLUSIONS: Guideline-based alert parameters in CIED patients significantly reduced NAA burden with no increasing in adverse outcomes in patients with device-detected AF or NSVT alerts. This approach may reduce noise and safely improve efficiency.
Предметные рубрики MeSH
HumansMaleFemaleAgedDefibrillators, ImplantableRisk FactorsPacemaker, ArtificialTime FactorsTreatment OutcomeAged, 80 and overDeath, Sudden, CardiacRisk AssessmentAtrial FibrillationMiddle AgedElectric CountershockClinical AlarmsIschemic StrokeTachycardia, VentricularIncidenceRetrospective StudiesRemote Sensing TechnologyHeart Rate