BACKGROUND

Detailed multidimensional assessment of patent foramen ovale (PFO) size with transesophageal echocardiography (TOE) may help to determine PFO pathogenicity in cryptogenic stroke patients. We explored the potential additive value of Live xPlane and three-dimensional (3D) TOE anatomical PFO sizing techniques.

METHODS

Imaging data of 45 patients who underwent a 3D TOE-assisted percutaneous PFO closure were studied. The two-dimensional (2D) PFO separation distance and right-to-left (RL) contrast shunt magnitude were assessed on preprocedural TOE recordings. Peri-procedural measurements of the triangular anatomical PFO opening (base, height, and area) were performed after positioning of a stiff guidewire (SW) through the PFO, using Live xPlane imaging and 3D Zoom mode.

RESULTS

The PFO SW base appeared on average 5 times larger than the preprocedural 2D PFO separation (median difference [IQR] = 13[5] mm; P < .001). For a same PFO separation, the width of the PFO base may vary significantly. The PFO SW base was significantly larger in patients with a large versus a small-to-moderate PFO RL contrast shunt (18 vs 15 mm; P = .007) and in those with a spontaneous versus a provoked shunt (18 vs 14 mm; P = .003).

CONCLUSION

Live xPlane and 3D Zoom TOE allow peri-procedural measurement of the largest dimension of a PFO, which is the PFO base. Patients with a large or spontaneous RL contrast shunt appear to have a larger PFO base. The anatomical PFO base dimension may be taken into account for optimization of device and patient selection strategies.

Для цитирования: Demulier Laurent, Paelinck Bernard Pierre, Coomans Ilse, Hemelsoet Dimitri, De Backer Julie, Campens Laurence, De Wolf Daniel. A new dimension in patent foramen ovale size estimation.. Echocardiography (Mount Kisco, N.Y.). 2020. DOI: 10.1111/echo.14696.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.