Flecainide causing elevated capture thresholds on pacemaker implantation.
Аннотация
Flecainide is a Vaughan Williams class 1c antiarrhythmic used to treat supraventricular and ventricular arrhythmias. It has been described as a rare cause for increased pacemaker capture thresholds. We describe a report of a patient, in her early 80s, presenting with tachy-brady syndrome on a background of permanent atrial fibrillation. She was treated with metoprolol and flecainide by her private cardiologist. Permanent right ventricular chamber pacing was recommended for her slow heart rate. At insertion of her single chamber pacemaker, she was noted to have elevated capture thresholds despite appropriate lead positioning. A flecainide level was elevated at 1.1 µg/mL, and it was subsequently ceased. This was associated with a rapid improvement in her capture threshold. Flecainide should be considered as a cause for elevated pacing thresholds at the time of implant. Particular care should be taken for at-risk groups such as the elderly and patients with renal impairment.