PloS one · 2022

Red and white blood cell morphology characterization and hands-on time analysis by the digital cell imaging analyzer DI-60.

СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Kweon Oh Joo, Lim Yong Kwan, Lee Mi-Kyung, Kim Hye Ryoun

Аннотация

BACKGROUND: The Sysmex DI-60 digital morphology analyzer is a fully automated, cell-locating image analysis system. This study aimed to evaluate the analytical performance of DI-60.

METHODS: A total of 822 peripheral blood smears were used. The diagnostic performance of DI-60 in terms of red blood cell (RBC) morphology characterization, white blood cell (WBC) differentials, and the total assay time including hands-on time was evaluated.

RESULTS: In comparison with manual slide review, DI-60 demonstrated acceptable accuracy in recognizing polychromasia, target cells, and ovalocytes. However, for schistocytes, DI-60 demonstrated low specificity (10.4%) despite the high sensitivity (97.2%). In the precision analysis of RBC morphology characterization, borderline samples harboring specific RBCs showed inconsistencies in the positive results among 20 replicates. Particularly, 6 of 10 samples showed inconsistencies in the precision for schistocytes. For WBC differentials, the overall agreement between pre-classification results and user-verified results was 89.4%. Except for basophils, normal WBCs showed a good correlation between DI-60 (after user verification) and manual counts. The sensitivities in detecting immature granulocytes, blasts, atypical lymphocytes, and normoblasts were 85.9%, 92.0%, 37.5%, and 77.6%, respectively. Although the total assay time of DI-60 was longer than that of manual review, the hands-on time was considerably shorter with a difference of 144.1 s/slide for abnormal samples.

CONCLUSION: DI-60 demonstrated acceptable performance for normal samples. However, for abnormal WBC differentials and RBC morphology characterization, it should be utilized carefully. DI-60 may contribute to an improvement in laboratory efficiency with increased feasibility.

Предметные рубрики MeSH