IEEE transactions on medical imaging · 2021
Development of an Integrated C-Arm Interventional Imaging System With a Strip Photon Counting Detector and a Flat Panel Detector.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Ji Xu, Feng Mang, Treb Kevin, Zhang Ran, Schafer Sebastian, Li Ke
Аннотация
Modern interventional x-ray systems are often equipped with flat-panel detector-based cone-beam CT (FPD-CBCT) to provide tomographic, volumetric, and high spatial resolution imaging of interventional devices, iodinated vessels, and other objects. The purpose of this work was to bring an interchangeable strip photon-counting detector (PCD) to C-arm systems to supplement (instead of retiring) the existing FPD-CBCT with a high quality, spectral, and affordable PCD-CT imaging option. With minimal modification to the existing C-arm, a 51×0.6 cm2 PCD with a 0.75 mm CdTe layer, two energy thresholds, and 0.1 mm pixels was integrated with a Siemens Artis Zee interventional imaging system. The PCD can be translated in and out of the field-of-view to allow the system to switch between FPD and PCD-CT imaging modes. A dedicated phantom and a new algorithm were developed to calibrate the projection geometry of the narrow-beam PCD-CT system and correct the gantry wobbling-induced geometric distortion artifacts. In addition, a detector response calibration procedure was performed for each PCD pixel using materials with known radiological pathlengths to address concentric artifacts in PCD-CT images. Both phantom and human cadaver experiments were performed at a high gantry rotation speed and clinically relevant radiation dose level to evaluate the spectral and non-spectral imaging performance of the prototype system. Results show that the PCD-CT system has excellent image quality with negligible artifacts after the proposed corrections. Compared with FPD-CBCT images acquired at the same dose level, PCD-CT images demonstrated a 53% reduction in noise variance and additional quantitative imaging capability.