MedElevator Войти в исследования
← Поиск публикаций

DOI 10.23750/abm.v94is2.13888

Медицинская библиотека · Acta bio-medica : Atenei Parmensis · 2023

Total hip arthroplasty monobloc stem neck rupture in patient with massive heterotopic ossifications: a case report and literature review.

Canton Gianluca, Barbo Giovanni, Roberto Fattori Roberto, Nicholas Rasio Nicholas, Murena Luigi

СтатьяПеревод в очереди

Аннотация

BACKGROUND AND AIM OF THE WORK: Fractures of the femoral stem neck are a rare complication in hip prosthetic surgery, especially in non-modular components. The authors report a case associated with massive heterotopic ossifications, with the purpose to analyze risk factors and specific characteristics.

METHODS

A case of femoral monobloc stem neck rupture is described. A non-systematic literature review regarding risk factors for femoral stem neck fracture was conducted in the PubMed database.

RESULTS

We report the case of a 61-year-old male who underwent surgery to remove calcifications four years after THA. Four months later the patient reported acute pain in the left hip, arising after a combined movement of external rotation and axial load while standing on the left foot, in the absence of any prodromic symptom. On radiographs, a displaced fracture of the neck of the hip prosthesis was revealed, together with massive heterotopic ossifications. After THA revision the patient's symptoms were resolved.

CONCLUSIONS

Prosthetic femoral neck fractures are a rare complication. We suggest that this case represents a unique type of fatigue rupture, where neck length and the presence of massive heterotopic calcifications contributed to flexion forces, resulting in failure in the midpoint of the neck.

Для цитирования: Canton Gianluca, Barbo Giovanni, Roberto Fattori Roberto, Nicholas Rasio Nicholas, Murena Luigi. Total hip arthroplasty monobloc stem neck rupture in patient with massive heterotopic ossifications: a case report and literature review.. Acta bio-medica : Atenei Parmensis. 2023. DOI: 10.23750/abm.v94is2.13888.