Clinical biomechanics (Bristol, Avon) · 2015

Angular stable lateral plating is a valid alternative to conventional plate fixation in the proximal phalanx. A biomechanical study.

СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Shanmugam R, Ernst M, Stoffel K, Fischer M F, Wahl D, Richards R G, Gueorguiev B

Аннотация

BACKGROUND: Dorsal plating is commonly used in proximal phalanx fractures but it bears the risk of interfering with the extensor apparatus. In this study, dorsal and lateral plating fixation methods are compared to assess biomechanical differences using conventional 1.5mm non-locking plates and novel 1.3mm lateral locking plates.

METHODS: Twenty-four fresh frozen human cadaveric proximal phalanges were equally divided into four groups. An osteotomy was set at the proximal metaphyseal-diaphyseal junction and fixed with either dorsal (group A) or lateral (group B) plating using a 1.5mm non-locking plate, or lateral plating with a novel 1.3mm locking plate with bicortical (group C) or unicortical (group D) screws. The specimens were loaded in axial, dorsovolar and mediolateral direction to assess fixation stiffness followed by a cyclic destructive test in dorsovolar loading direction.

FINDINGS: Axial stiffness was highest in group D (mean 321.02, SEM 21.47N/mm) with a significant difference between groups D and B (P=0.033). Locking plates (groups C and D) were stiffer than non-locking plates under mediolateral loading (P=0.007), no significant differences were noted under dorsovolar loading. Furthermore, no significant differences were observed under cyclic loading to failure between any of the study groups.

INTERPRETATION: No considerable biomechanical advantage of using a conventional 1.5mm dorsal non-locking plate was identified over the novel 1.3mm lateral locking plate in the treatment of proximal phalanx fractures. Since the novel low-profile plate is less disruptive to the extensor mechanism, it should be considered as a valid alternative.