[Sprint fidelis defibrillation lead: a nine-center experience in Spain].

Arias Miguel A, Domínguez-Pérez Laura, Toquero Jorge, Jiménez-Candil Javier, Olagüe José, Díaz-Infante Ernesto, Tercedor Luis, Valverde Irene, Castro Jorge, García-Fernández Francisco J, Rodríguez-Padial Luis

Аннотация на английскомСтатьяМногоцентровое исследованиеПеревод в очереди

Аннотация

INTRODUCTION AND

OBJECTIVES

Sprint Fidelis defibrillation leads are prone to early failure. Most of the reported series come from a single institution. This paper describes the clinical experience in nine Spanish hospitals.

METHODS

Clinical, implant, and follow-up data of all patients with a Sprint Fidelis lead were analyzed. All cases of lead failure were identified, medium-term lead survival was calculated, and possible predictors for lead failure were determined.

RESULTS

In total, 378 leads in 376 patients were studied. The mean age (male 85.7%) was 64.9 ± 13.6 years. The majority of patients (59.8%) had ischemic heart disease. Mean left ventricular ejection fraction was 33.4% ± 14.5%. Left subclavian vein puncture was used in 74.8%. During a mean follow-up of 30.9 ± 14 months, 16 lead failures have occurred, with a lead survival of 96.1% at 36 months after implantation. Eleven of 16 lead failures were caused by failure of pace/sense conductors, 3 by defects in the high-voltage conductor, and 2 by defects in both types of conductors. A less depressed left ventricular ejection fraction was associated with an increased probability of lead failure (42.4% ± 16% vs. 33% ± 14.3%; P =.011). Three hospitals presented a rate of lead failure higher than 10%; the rate was less than 5% in the remaining 6 hospitals.

CONCLUSIONS

In this multicenter series of 378 leads, the 3-year estimated survival was higher than that reported in prior series. Clinical presentation of lead failures was similar to that reported previously. Left ventricular ejection fraction and hospital of implantation were variables associated to lead failure.

Для цитирования: Arias Miguel A, Domínguez-Pérez Laura, Toquero Jorge, Jiménez-Candil Javier, Olagüe José, Díaz-Infante Ernesto, Tercedor Luis, Valverde Irene, Castro Jorge, García-Fernández Francisco J, Rodríguez-Padial Luis. [Sprint fidelis defibrillation lead: a nine-center experience in Spain].. Revista espanola de cardiologia. 2011. DOI: 10.1016/j.recesp.2010.11.008.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.