DOI 10.1016/j.cnp.2018.06.003

Медицинская библиотека · Clinical neurophysiology practice · 2018

Evaluation of SEPs in asphyxiated newborns using a 4-electrode aEEG brain monitoring set-up.

Nevalainen Päivi, Marchi Viviana, Metsäranta Marjo, Lönnqvist Tuula, Vanhatalo Sampsa, Lauronen Leena

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

To evaluate the reliability of recording cortical somatosensory evoked potentials (SEPs) in asphyxiated newborns using the 4-electrode setup applied in routine long-term amplitude-integrated EEG (aEEG) brain monitoring and to assess the number of averages needed for reliably detecting the cortical responses.

METHODS

We evaluated median nerve SEPs in 50 asphyxiated full-term newborns. The SEP interpretation (present or absent) from the original recordings with 21-electrodes and approximately 600 trials served as the reference. This was compared to SEP classification (absent, present, or unreliable) based on a reduced (300 or 150) number of averages, and to classification based on only four electrodes (F3, P3, F4, P4).

RESULTS

Compared to the original classification, cortical SEPs were uniformly interpreted as present or absent in all 50 newborns with the 4-electrode setup and 600 averages. Reducing number of averages to 300 still resulted in correct SEP interpretation in 49/50 newborns with 21-electrode setup, and 46/50 newborns with 4-electrode setup.

CONCLUSIONS

Evaluation of early cortical neonatal SEPs is reliable from the 4-electrode setup commonly used in aEEG monitoring. SEP is discernible in most newborns with 300 averages.

SIGNIFICANCE: Adding SEP into routine aEEG monitoring offers an additional tool for early neonatal neurophysiological evaluation.

Для цитирования: Nevalainen Päivi, Marchi Viviana, Metsäranta Marjo, Lönnqvist Tuula, Vanhatalo Sampsa, Lauronen Leena. Evaluation of SEPs in asphyxiated newborns using a 4-electrode aEEG brain monitoring set-up.. Clinical neurophysiology practice. 2018. DOI: 10.1016/j.cnp.2018.06.003.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.