Kelkar Neil, Eid Kevin, Nakatsuka Austin S, Johnson Catherine, Mamalis Nick, Werner Liliana

СтатьяПоддержка NIH: внешняя программаПоддержка службы здравоохранения СШАПеревод в очереди

Аннотация

PURPOSE: To evaluate CT Lucia 602 intraocular lenses implanted in cases lacking capsular support and explanted because of decentration/dislocation with subsequent pathological analysis. The main objectives were to (1) document the status of haptics and (2) verify the presence of the rotisserie effect.

SETTING

John A. Moran Eye Center, University of Utah, Salt Lake City, Utah.

DESIGN

Case series with clinicopathological correlation.

METHODS

13 explanted CT Lucia 602 lenses were included: 12 implanted by the Yamane technique and 1 that was iris-fixated. 4 lenses were explanted between October 2020 and February 2022. 9 lenses, explanted between November 2022 and February 2024, also exhibited severe optic tilt, leading to explantation. A chart review was performed to obtain clinical information from each case. Explanted lenses underwent gross and microscopic examination for haptic status and rotation, in the dry state (room temperature) and after hydration (body temperature).

RESULTS

Time between implantation and explantation was longer for lenses explanted before November 2022 (without severe tilt). Among the 26 haptics evaluated, 15 were deformed, 6 were broken, 2 were detached, and 3 were unremarkable. 4 haptics of Yamane lenses explanted between November 2022 and February 2024 showed free rotation within the optic insertion, after hydration at body temperature.

CONCLUSIONS

Anecdotal reports of the rotisserie effect of CT Lucia 602 lenses implanted by the Yamane technique appeared mostly during the second half of 2022 and suggested that affected lenses were from specific batches, which seems to be consistent with the findings of this study.

Для цитирования: Kelkar Neil, Eid Kevin, Nakatsuka Austin S, Johnson Catherine, Mamalis Nick, Werner Liliana. Severe intraocular lens tilt after the Yamane technique.. Journal of cataract and refractive surgery. 2025. DOI: 10.1097/j.jcrs.0000000000001575.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.