Аннотация

BACKGROUND

Intraoperative surgical navigation has become widely accepted as an important tool for improvement of surgical outcomes and reduction of complication in endoscopic sinus surgery (ESS). The purpose of this study was to assess the clinical utility of intraoperative volume computed tomography (CT) scanning in endoscopic sinonasal and skull base procedures.

METHODS

Retrospective review of patients who underwent intraoperative volume CT imaging (xCAT; XoranTechnologies, Ann Arbor, MI) during endoscopic sinonasal and skull base surgery during a 3-month period was performed. Intraoperative, computer-enabled triplanar review of reformatted 0.4-mm images was performed in all cases.

RESULTS

Intraoperative volume CT scanning was completed in 25 patients. Surgical procedures included revision/primary ESS for chronic rhinosinusitis (CRS) with or without polyposis (12 cases) and mucoceles (6 cases) as well as endoscopic neoplasm resection (5 cases), endoscopic fibro-osseous lesion resection (1 case), and endoscopic meningoencephalocele repair (1 case). The indications for intraoperative CT scanning included assessment of surgical dissection (23 cases), extent of tumor resection (6 cases), and frontal stent placement (6 cases). Based on the intraoperative volume CT information, additional interventions, including additional tumor resection (2 cases), dissection of ethmoid partitions (2 cases), frontal bone drilling during Draf IIB (1 case), and repositioning of a frontal stent (1 case) were performed in 6 (24%) cases.

CONCLUSION

Intraoperative volume CT scanning was successfully performed in 25 patients undergoing ESS. Because additional surgical intervention was performed in 24% of cases, this technology may have an important role in endoscopic sinonasal and skull base procedures.

Для цитирования: Batra Pete S, Kanowitz Seth J, Citardi Martin J. Clinical utility of intraoperative volume computed tomography scanner for endoscopic sinonasal and skull base procedures.. American journal of rhinology. 2008. DOI: 10.2500/ajr.2008.22.3216.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.