MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.cardfail.2023.04.007

Медицинская библиотека · Journal of cardiac failure · 2023

Modification of a Transvalvular Microaxial Flow Pump for Instantaneous Determination of Native Cardiac Output and Volume.

Aslam M Imran, Gruslova Aleksandra B, Almomani Ahmed, Nolen Drew, Elliott James J, Jani Vivek P, Kottam Anil, Porterfield John, Heighten Clay, Anderson Allen S, Valvano Jonathan W, Feldman Marc D

СтатьяПоддержка исследования: вне правительства СШАПоддержка NIH: внешняя программаПеревод в очереди

Аннотация

BACKGROUND

The current Impella cardiopulmonary (CP) pump, used for mechanical circulatory support in patients with cardiogenic shock (CS), cannot assess native cardiac output (CO) and left ventricular (LV) volumes. These data are valuable in facilitating device management and weaning. Admittance technology allows for accurate assessment of cardiac chamber volumes.

OBJECTIVES

This study tested the ability to engineer admittance electrodes onto an existing Impella CP pump to assess total and native CO as well as LV chamber volumes in an instantaneous manner.

METHODS

Impella CP pumps were fitted with 4 admittance electrodes and were placed in the LVs of adult swine (n = 9) that were subjected to 3 different hemodynamic conditions, including Impella CP speed adjustments, administration of escalating doses of dobutamine and microsphere injections into the left main artery to result in cardiac injury. CO, according to admittance electrodes, was calculated from LV volumes and heart rate. In addition, CO was calculated in each instance via thermodilution, continuous CO measurement, the Fick principle, and aortic velocity-time integral by means of echocardiography.

RESULTS

Modified Impella CP pumps were placed in swine LVs successfully. CO, as determined by admittance electrodes, was similar by trend to other methods of CO assessment. It was corrected for pump speed to calculate native CO, and calculated LV chamber volumes trended as expected in each experimental protocol.

CONCLUSIONS

We report, for the first time, that an Impella CP pump can be fitted with admittance electrodes and used to determine total and native CO in various hemodynamic situations.

CONDENSED ABSTRACT: Transvalvular mechanical circulatory support devices such as the Impella CP do not have the ability to provide real-time information on native cardiac output (CO) and left ventricular (LV) volumes. This information is critical in device management and in weaning in patients with cardiogenic shock. We demonstrate, for the first time, that Impella CP pumps coupled with admittance electrodes are able to determine native CO and LV chamber volumes in multiple hemodynamic situations such as Impella pump speed adjustments, escalating dobutamine administration and cardiac injury from microsphere injection.

Для цитирования: Aslam M Imran, Gruslova Aleksandra B, Almomani Ahmed, Nolen Drew, Elliott James J, Jani Vivek P, Kottam Anil, Porterfield John, Heighten Clay, Anderson Allen S, Valvano Jonathan W, Feldman Marc D. Modification of a Transvalvular Microaxial Flow Pump for Instantaneous Determination of Native Cardiac Output and Volume.. Journal of cardiac failure. 2023. DOI: 10.1016/j.cardfail.2023.04.007.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.