An observational study to quantify manual adjustments of the inspired oxygen fraction in extremely low birth weight infants.

Оценочное исследованиеСтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

van der Eijk Anne Catherine, Dankelman Jenny, Schutte Sander, Simonsz Huibert Jan, Smit Bert Johan

Аннотация

AIM: To quantify manual fraction of inspired oxygen (FiO(2)) adjustments performed by caregivers in extremely low birth weight (ELBW; ≤1000 g) infants, in relation to oxygen saturation (SpO(2)) and bedside care.

METHODS: In a single-centre study, FiO(2) , SpO(2) and alarm limits of ELBW infants were collected for 3 days continuously, while caregivers were filmed. A descriptive analysis, focused on manual FiO(2) adjustments, was performed.

RESULTS: Twelve ELWB infants were included. Total recording time was 726 h. FiO(2) was increased 851 times and decreased 1309 times; median (range) step size was 5% (1% to 65%) and -3% (-1% to -65%), respectively. Wide variation of FiO(2) adjustments for equal levels of SpO(2) was observed in all included infants. One hundred and twenty-six of 136 FiO(2) adjustments with a step size ≥15% and 111 of 171 desaturations <70% were associated with medical or nursing procedures. When FiO(2) was >21%, alarm limits for SpO(2) were set according to protocol (88-94%) in 64% of the time. Within these periods, SpO(2) was >94% for 30% and <88% for 16% of the time.

CONCLUSIONS: Manual FiO(2) adjustments varied widely in frequency and step size. Deep desaturations and large FiO(2) adjustments were associated with medical or nursing procedures. When large adjustments are really necessary, it will be challenging to implement them in an automatic adjustment device.