Continuous monitoring of the partial pressure of oxygen in cerebral venous blood.

Rieger A, Menzel M, Rainov N G, Sanchin L, Rot S, Furka I, Görömbey Z, Burkert W

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

OBJECTIVE

Clinical oxygen monitoring in the injured brain is somewhat difficult. However, ischemia is one of the major factors responsible for secondary tissue damage after head injury or subarachnoid hemorrhage. Therefore, the aim of the present study was to investigate the value of continuously monitoring the partial pressure of oxygen in cerebral venous blood (PcvO2) during changes in intracranial pressure (ICP).

METHODS

In eight domestic pigs with Clark type probes placed in the posterior third of the superior sagittal sinus, PcvO2 was continuously registered while ICP was stepwise elevated by an inflatable balloon placed below the tentorium. Arterial blood pressure was continuously monitored, cerebral perfusion pressure (CPP) was calculated, and arterial partial carbon dioxide pressure and partial pressure of oxygen were registered intermittently.

RESULTS

The mean intraparenchymal ICP before the start of balloon inflation was 5 +/- 1 mm Hg, the mean CPP was 80 +/- 15 mm Hg, and the mean PcvO2 was 36 +/- 3 mm Hg. At maximum ICP elevation, CPP decreased to 20 +/- 12 mm Hg, PcvO2 decreased to 10 +/- 6 mm Hg, and ICP increased to 90 +/- 10 mm Hg. Strong linear correlations between ICP and PcvO2 and between CPP and PcvO2 were revealed, and mean correlation coefficients of 0.89 for ICP/PcvO2 and 0.73 for CPP/PcvO2 were calculated.

CONCLUSION

The present study demonstrates that polarographic PcvO2 monitoring in the superior sagittal sinus is a reliable method for the early detection of reduced CPP during ICP elevation. This technique is capable of registering the global oxygen supply and oxygen consumption of the brain. It seems superior to jugular venous oxymetry and is better suited for clinical use because of a somewhat low artifact susceptibility.

Для цитирования: Rieger A, Menzel M, Rainov N G, Sanchin L, Rot S, Furka I, Görömbey Z, Burkert W. Continuous monitoring of the partial pressure of oxygen in cerebral venous blood.. Neurosurgery. 1997. DOI: 10.1097/00006123-199708000-00027.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.