Oscillometric estimation of central blood pressure: validation of the Mobil-O-Graph in comparison with the SphygmoCor device.
Сравнительное исследованиеСтатьяВалидационное исследованиеНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Weiss Wolfgang, Gohlisch Christopher, Harsch-Gladisch Christl, Tölle Markus, Zidek Walter, van der Giet Markus
Аннотация
BACKGROUND: Hypertension is a major risk factor for a wide range of cardiovascular diseases and is typically identified by measuring blood pressure (BP) at the brachial artery. Although such a measurement may accurately determine diastolic BP, systolic BP is not reflected accurately. Current noninvasive techniques for assessing central aortic BP require additional recording of an arterial pressure wave using a high-fidelity applanation tonometer. Within one measurement cycle, the Mobil-O-Graph BP device uses brachial oscillometric BP waves for a noninvasive estimation of central BP. We therefore validated the Mobil-O-Graph against the SphygmoCor device, which is widely known as the commonly used approach for a noninvasive estimation of central BP.
METHODS: For each individual, we compared three readings of the central BP values obtained by the Mobil-O-Graph and SphygmoCor device consecutively. One hundred individuals (mean age 56.1 ± 15.4 years) were recruited for measurement.Differences between the central BP values of the test device and the SphygmoCor device were calculated for each measurement.
RESULTS: The mean difference (95% confidence interval) for the estimated central systolic BP between both devices was -0.6 ± 3.7 mmHg. Comparison of the central BP values measured by the two devices showed a statistically significant linear correlation (R=0.91, P<0.0001). The mean between-method difference was 0.50 mmHg for central systolic BP estimation. The intrarater reproducibility between both the devices was also comparable. Bland and Altman analyses showed that the mean differences (95% confidence interval) between repeated measurements were 1.89 (0.42-3.36) mmHg and 1.36 (-0.16 to 2.83) mmHg for the SphygmoCor and the Mobil-O-Graph device, respectively. Thus, neither of these differences was statistically significantly different from 0. The limits of agreement were -16.34 to 19.73 and -15.23 to 17.17 mmHg for the SphygmoCor and the Mobil-O-Graph device, respectively.
CONCLUSION: Oscillometric noninvasive estimation of central BP with the Mobil-O-Graph BP device is as effective as using the well-established SphygmoCor applanation tonometry device. In comparison, the Mobil-O-Graph combines the widespread benefits of brachial BP measurement and also provides central BP within one measurement.
METHODS: For each individual, we compared three readings of the central BP values obtained by the Mobil-O-Graph and SphygmoCor device consecutively. One hundred individuals (mean age 56.1 ± 15.4 years) were recruited for measurement.Differences between the central BP values of the test device and the SphygmoCor device were calculated for each measurement.
RESULTS: The mean difference (95% confidence interval) for the estimated central systolic BP between both devices was -0.6 ± 3.7 mmHg. Comparison of the central BP values measured by the two devices showed a statistically significant linear correlation (R=0.91, P<0.0001). The mean between-method difference was 0.50 mmHg for central systolic BP estimation. The intrarater reproducibility between both the devices was also comparable. Bland and Altman analyses showed that the mean differences (95% confidence interval) between repeated measurements were 1.89 (0.42-3.36) mmHg and 1.36 (-0.16 to 2.83) mmHg for the SphygmoCor and the Mobil-O-Graph device, respectively. Thus, neither of these differences was statistically significantly different from 0. The limits of agreement were -16.34 to 19.73 and -15.23 to 17.17 mmHg for the SphygmoCor and the Mobil-O-Graph device, respectively.
CONCLUSION: Oscillometric noninvasive estimation of central BP with the Mobil-O-Graph BP device is as effective as using the well-established SphygmoCor applanation tonometry device. In comparison, the Mobil-O-Graph combines the widespread benefits of brachial BP measurement and also provides central BP within one measurement.