BACKGROUND

Attenuation of the peak impact force is essential in any protective devices for prevention of fall-related injuries.

HYPOTHESIS: Common wrist guards have limited effectiveness because of the multifaceted nature of wrist injury mechanisms, and other modalities may provide enhanced shock-absorbing functions.

STUDY

DESIGN

Controlled laboratory study.

METHODS

A free-fall device was constructed using a mechanical surrogate to simulate falling impact. At 4 different falling heights, 5 different hand conditions were tested: bare hand, a generic-brand wrist guard, a Sorbothane glove, an air cell, and an air bladder condition. The impact force from the ground and the transmitted impact force to the forearm/hand complex were simultaneously measured.

RESULTS

The falling height and hand condition significantly modulated the impact responses. The padded conditions always had significantly smaller peak impact forces compared with the bare-hand condition. The wrist guard became ineffective in impact force attenuation beyond the falling height of 51 cm. On the other hand, the air bladder condition maintained less than 45% of the peak impact force of the bare-hand condition and remained below the critical value, whereas other conditions were all ineffective.

CONCLUSION

It was reconfirmed that common wrist guard design could provide limited impact force attenuation, whereas damped pneumatic springs would provide substantially enhanced shock-absorbing functions.

CLINICAL RELEVANCE: A wrist guard incorporating volar padding with the pneumatic spring design principle might be more effective at preventing injuries than are currently available designs.

Для цитирования: Kim Kyu-Jung, Alian Ali M, Morris William S, Lee Young-Hwa. Shock attenuation of various protective devices for prevention of fall-related injuries of the forearm/hand complex.. The American journal of sports medicine. 2005. DOI: 10.1177/0363546505281800.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.