Validation method for determining the concentration of norfloxacin-tylosin combination in broiler chicken tissue using high-performance liquid chromatography.
СтатьяВалидационное исследованиеНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Wibisono Cahyo, Wijayanti Agustina Dwi, Widiasih Dyah Ayu, Tjahajati Ida, Irianti Ika Nindya, Pamudya Kanya Gabriella Sasri, Salsabila Ridha Avicena Ila, Sudimartini Luh Made
Аннотация
BACKGROUND: The combination of norfloxacin and tylosin is an alternative to prevent antibiotic residue and resistance. Measurement of the levels of the combination of norfloxacin-tylosin in tissue using high-performance liquid chromatography (HPLC) to obtain parameters for further research, such as pharmacokinetics and residue analysis. When broiler chickens are given norfloxacin-tylosin antibiotics, they make antibiotics more effective against bacterial illnesses. These results indicate that the right mix of antibiotics can be used as a single medicine to achieve synergistic effects. Also, reach the SDGs for no hunger, good health, and well-being.
AIM: This study aimed to ascertain the correct validation method for the combination antibiotic norfloxacin-tylosin in broiler chicken tissues, specifically the liver, kidney, and muscle.
METHODS: HPLC consistently detected norfloxacin-tylosin in broiler chicken tissue. Mobile phase methanol/acetonitrile was 50:50 (v/v). The chromatographic equipment utilized a 250 mm Shimpack ODS C18 column with a four μm diameter. Solution detection happened at a 1 ml/minutes flow rate, 30°C oven temperature, and 20 μl injection volume. Light had a wavelength of 279 nm. In Bogor, Indonesia, PT Mensana Aneka Satwa manufactured liquid antibiotics and combined them with broiler chicken tissue from 28-day-old antibiotic-free-fed chicks.
RESULTS: The tissue, liver, kidney, and muscle showed linearity values of R 2 = 0.9999, R 2 = 0.9992, and R 2 = 0.9939 for norfloxacin and R 2 = 0.9969, R 2 = 0.9964, and R 2 = 0.9938 for tylosin. Norfloxacin had liver, kidney, and muscle accuracy scores of 100.57%, 101.85%, and 105.41%, respectively, while tylosin had 103.89%, 95.83%, and 105.49%. Norfloxacin showed intraday precision values (%CV) of 0.77%, 4.18%, and 4% in the liver, kidney, and muscle, while tylosin had 0.74%, 1.84%, and 1%. Tylosin retained 3,325-3,358 minutes and norfloxacin 3.083-3.092. The limit of detection values for norfloxacin in the liver, kidney, and muscle were 1.57, 5.10, and 1.4 μg/ml, respectively. Tylosin levels in liver, kidney, and muscle were 1.72, 1.15, and 1.51 μg/ml, while norfloxacin had limit of quantification values of 5.24, 17.01, and 4.98 μg/ml.
CONCLUSION: According to the European Medicines Agency Guideline, the mixed solution of norfloxacin and tylosin demonstrated favourable outcomes regarding sensitivity, specificity, linearity, accuracy, and precision.
AIM: This study aimed to ascertain the correct validation method for the combination antibiotic norfloxacin-tylosin in broiler chicken tissues, specifically the liver, kidney, and muscle.
METHODS: HPLC consistently detected norfloxacin-tylosin in broiler chicken tissue. Mobile phase methanol/acetonitrile was 50:50 (v/v). The chromatographic equipment utilized a 250 mm Shimpack ODS C18 column with a four μm diameter. Solution detection happened at a 1 ml/minutes flow rate, 30°C oven temperature, and 20 μl injection volume. Light had a wavelength of 279 nm. In Bogor, Indonesia, PT Mensana Aneka Satwa manufactured liquid antibiotics and combined them with broiler chicken tissue from 28-day-old antibiotic-free-fed chicks.
RESULTS: The tissue, liver, kidney, and muscle showed linearity values of R 2 = 0.9999, R 2 = 0.9992, and R 2 = 0.9939 for norfloxacin and R 2 = 0.9969, R 2 = 0.9964, and R 2 = 0.9938 for tylosin. Norfloxacin had liver, kidney, and muscle accuracy scores of 100.57%, 101.85%, and 105.41%, respectively, while tylosin had 103.89%, 95.83%, and 105.49%. Norfloxacin showed intraday precision values (%CV) of 0.77%, 4.18%, and 4% in the liver, kidney, and muscle, while tylosin had 0.74%, 1.84%, and 1%. Tylosin retained 3,325-3,358 minutes and norfloxacin 3.083-3.092. The limit of detection values for norfloxacin in the liver, kidney, and muscle were 1.57, 5.10, and 1.4 μg/ml, respectively. Tylosin levels in liver, kidney, and muscle were 1.72, 1.15, and 1.51 μg/ml, while norfloxacin had limit of quantification values of 5.24, 17.01, and 4.98 μg/ml.
CONCLUSION: According to the European Medicines Agency Guideline, the mixed solution of norfloxacin and tylosin demonstrated favourable outcomes regarding sensitivity, specificity, linearity, accuracy, and precision.