MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1136/hrt.64.6.393

Медицинская библиотека · British heart journal · 1990

Patients' observations of bioprosthetic valve failure: "my heart is honking, doctor".

Errington M, Bloomfield P, Starkey I R, Shaw T R

Клинический случайСтатьяПеревод в очереди

Аннотация

Audible cardiac murmurs are uncommon and may indicate severe native valve dysfunction. In six patients with bioprosthetic mitral valves audible honking systolic murmurs suddenly developed. Doppler echocardiography showed characteristic striated regurgitant signals. At operation each prosthesis had torn cusps. New audible murmurs and striated Doppler signals originating from mitral bioprosthetic valves are indicators of cusp tears. Recognition of this is important because early detection of prosthetic valve dysfunction may improve subsequent surgical outcome.

Для цитирования: Errington M, Bloomfield P, Starkey I R, Shaw T R. Patients' observations of bioprosthetic valve failure: "my heart is honking, doctor".. British heart journal. 1990. DOI: 10.1136/hrt.64.6.393.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.