DOI 10.1186/1471-2474-13-32

Медицинская библиотека · BMC musculoskeletal disorders · 2012

Osseointegration and biocompatibility of different metal implants--a comparative experimental investigation in sheep.

Plecko Michael, Sievert Christine, Andermatt Daniel, Frigg Robert, Kronen Peter, Klein Karina, Stübinger Stefan, Nuss Katja, Bürki Alexander, Ferguson Stephen, Stoeckle Ulrich, von Rechenberg Brigitte

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

In the present study, 4 different metallic implant materials, either partly coated or polished, were tested for their osseointegration and biocompatibility in a pelvic implantation model in sheep.

METHODS

Materials to be evaluated were: Cobalt-Chrome (CC), Cobalt-Chrome/Titanium coating (CCTC), Cobalt-Chrome/Zirconium/Titanium coating (CCZTC), Pure Titanium Standard (PTST), Steel, TAN Standard (TANST) and TAN new finish (TANNEW). Surgery was performed on 7 sheep, with 18 implants per sheep, for a total of 63 implants. After 8 weeks, the specimens were harvested and evaluated macroscopically, radiologically, biomechanically (removal torque), histomorphometrically and histologically.

RESULTS

Cobalt-Chrome screws showed significantly (p = 0.031) lower removal torque values than pure titanium screws and also a tendency towards lower values compared to the other materials, except for steel. Steel screws showed no significant differences, in comparison to cobalt-chrome and TANST, however also a trend towards lower torque values than the remaining materials. The results of the fluorescence sections agreed with those of the biomechanical test. Histomorphometrically, there were no significant differences of bone area between the groups. The BIC (bone-to-implant-contact), used for the assessment of the osseointegration, was significantly lower for cobalt-chrome, compared to steel (p = 0.001). Steel again showed a lower ratio (p = 0.0001) compared to the other materials.

CONCLUSION

This study demonstrated that cobalt-chrome and steel show less osseointegration than the other metals and metal-alloys. However, osseointegration of cobalt-chrome was improved by zirconium and/or titanium based coatings (CCTC, TANST, TAN, TANNEW) being similar as pure titanium in their osseointegrative behavior.

Для цитирования: Plecko Michael, Sievert Christine, Andermatt Daniel, Frigg Robert, Kronen Peter, Klein Karina, Stübinger Stefan, Nuss Katja, Bürki Alexander, Ferguson Stephen, Stoeckle Ulrich, von Rechenberg Brigitte. Osseointegration and biocompatibility of different metal implants--a comparative experimental investigation in sheep.. BMC musculoskeletal disorders. 2012. DOI: 10.1186/1471-2474-13-32.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.