Transaxillary transcatheter aortic valve implantation utilizing a novel vascular closure device with resorbable collagen material: a feasibility study.
Аннотация
OBJECTIVES
We herein aimed to evaluate technical feasibility of transaxillary (Tax) transcatheter aortic valve implantation (TAVI) utilizing a novel vascular closure device with a resorbable collagen plug and absence of suture material.
METHODS
Between 05/2018 and 8/2018, eight patients (76.0 ± 5.9 years, 62.5% male, logEuroSCORE I 23.6 ± 4.7) received Tax-TAVI using the MANTA™ vascular closure device. Implanted transcatheter heart valves consisted of Edwards Sapien 3, NVT Allegra, Medtronic CoreValve EvolutR and SJM Portico.
RESULTS
Puncture location depth was variable (3.5-7.5 cm). The left subclavian artery was used in five cases, the right subclavian artery in three cases. Low-pressure balloon-angioplasty for vessel closure was performed in 5/8 patients. VARC-2 defined device success was met in all patients. Major access site complication occurred in one patient with aneurysma spurium of the subclavian artery and consecutive stent implantation on postoperative day 5.
CONCLUSION
The MANTA™ device is applicable in Tax-TAVI, with potential particular advantages regarding easiness of use and marked access for subsequent interventions in case of vascular complications. Before conclusions regarding clinical efficacy and safety can be made, the device has to be evaluated in larger patient cohorts.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.