Thirty-minute compared to standardised office blood pressure measurement in general practice.

Scherpbier-de Haan Nynke, van der Wel Mark, Schoenmakers Gijs, Boudewijns Steve, Peer Petronella, van Weel Chris, Thien Theo, Bakx Carel

Сравнительное исследованиеОценочное исследованиеСтатьяПеревод в очереди

Аннотация

BACKGROUND

Although blood pressure measurement is one of the most frequently performed measurements in clinical practice, there are concerns about its reliability. Serial, automated oscillometric blood pressure measurement has the potential to reduce measurement bias and white-coat effect'.

AIM: To study agreement of 30-minute office blood pressure measurement (OBPM) with standardised OBPM, and to compare repeatability.

DESIGN AND

SETTING

Method comparison study in two general practices in The Netherlands.

METHOD: Thirty-minute and standardised OBPM was carried out with the same, validated device in 83 adult patients, and the procedure was repeated after 2 weeks. During 30-minute OBPM, blood pressure was measured automatically every 3 minutes, with the patient in a sitting position, alone in a quiet room. Agreement between 30-minute and standardised OBPM was assessed by Bland-Altman analysis. Repeatability of the blood pressure measurement methods after 2 weeks was expressed as the mean difference in combination with the standard deviation of difference (SDD).

RESULTS

Mean 30-minute OBPM readings were 7.6/2.5 mmHg (95% confidence interval [CI] = 6.1 to 9.1/1.5 to 3.4 mmHg) lower than standardised OBPM readings. The mean difference and SDD between repeated 30-minute OBPMs (mean difference = 3/1 mmHg, 95% CI = 1 to 5/0 to 2 mmHg; SDD 9.5/5.3 mmHg) were lower than those of standardised OBPMs (mean difference = 6/2 mmHg, 95% CI = 4 to 8/1 to 4 mmHg; SDD 10.9/6.3 mmHg).

CONCLUSION

Thirty-minute OBPM resulted in lower readings than standardised OBPM and had a better repeatability. These results suggest that 30-minute OBPM better reflects the patient's true blood pressure than standardised OBPM does.

Для цитирования: Scherpbier-de Haan Nynke, van der Wel Mark, Schoenmakers Gijs, Boudewijns Steve, Peer Petronella, van Weel Chris, Thien Theo, Bakx Carel. Thirty-minute compared to standardised office blood pressure measurement in general practice.. The British journal of general practice : the journal of the Royal College of General Practitioners. 2011. DOI: 10.3399/bjgp11x593875.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.