Biomechanical comparison of osteoporotic distal radius fractures fixed by distal locking screws with different length.

Liu Xiong, Wu Wei-dong, Fang Ya-feng, Zhang Mei-chao, Huang Wen-hua

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

OBJECTIVES

To evaluate the postoperative stability of osteoporotic distal radius fractures fixed with distal locking screws with different length.

METHODS

A comminuted extra-articular dorsally unstable distal radius fracture, treated with volar locking plate system, was created. The 18 specimens were randomized into 3 groups based on distal locked screws with different length: Group A had unicortical screws with 50% length to the dorsal cortex. Group B had unicortical screws with 75% length to the dorsal cortex. Group C had bicortical screws. Axial compression and bending loads were imposed on the models before and after cycling testing as well as load to clinical and catastrophic failure.

RESULTS

Minimum change in stiffness was observed before and after fatigue for all groups. The final stiffness to bending forces was statistically similar in all groups, but stiffness to axial compression was statistically significant different: Group A approached significance with respect to groups B and C (P = 0.017, 0.009), whereas stiffness in group B and C was statistically similar (P = 0.93). Load to clinical failure was significantly less for group A (456.54±78.59 N) compared with groups B (580.24±73.85 N) and C (591.07±38.40 N). Load to catastrophic failure was statistically similar between groups, but mean values for Group A were 18% less than means for Group C.

CONCLUSIONS

The volar locking plate system fixed with unicortical locking screws with at least 75% length not only produced early stability for osteoporotic distal radius fractures, but also avoided extensor tendon complications due to dorsal screw protrusion.

Для цитирования: Liu Xiong, Wu Wei-dong, Fang Ya-feng, Zhang Mei-chao, Huang Wen-hua. Biomechanical comparison of osteoporotic distal radius fractures fixed by distal locking screws with different length.. PloS one. 2014. DOI: 10.1371/journal.pone.0103371.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.